15.2
General Trouble Shooting
Symptom
ON/OFF button pressed to ON,
Joystick lights do not light up wheelchair
does not move.
ON/OFF lights up, wheelchair does not
move or behaves erratically.
Battery Charge indicator shows low
battery. Wheelchair moves slowly and
does not respond well to joystick
movement.
Batteries will not accept charge.
15.3
Shark Remote Information Gauge
Horn Button
ON/OFF Button
Replace remote
Power module
defective
Check / replace
batteries
Check / Action
Are the battery cables connected ?
Is the joystick cable connected ?
Are the batteries disconnected ?
Are the batteries completely flat ?
Is the fuse / circuit breaker open ?
Are one or both motors disengaged
(freewheel mode) ?
Is the Shark remote in Lock Mode? (See
user guide).
Are the batteries discharged to the
critical level ?
Possible charger fault ?
Batteries are no longer serviceable ?
Information Gauge
Speed Dial
Section 12.3
Section 12.3
Section 13.2
The following table gives the basic
status of the Shark remote.
Refer to the Information Gauge
for the relevant flash code or
status.