Troubleshooting

The power does not turn on
Cause
If the unit is being powered by the
batteries, the batteries are not
installed or depleted.
The AC adapter is disconnected.
If the power does not turn on and the above is not the cause, turn off the
unit power, disconnect the AC adapter, remove the batteries, and contact
your local OMRON representative.
The unit display does not operate
Cause / solution
Stop using the unit and contact your local OMRON representative.
The unit becomes hot
Cause
An object is on top of the unit or right
next to the unit.
If the unit becomes too hot to be touched, there may be a problem in the
unit. Turn off the unit power, disconnect the AC adapter, remove the
batteries, and contact your local OMRON representative.
The cuff does not inflate when the [START/STOP] button is pressed
Cause
Loose cuff tube connection.
There is an air leak in the cuff.
If pressure is displayed, the cuff tube
is bent.
26
Solution
Insert batteries or replace with new
batteries (page 13).
Connect the AC adapter (page 13).
Solution
Keep the area around the unit free of
objects.
Solution
Check the connection.
Replace the cuff.
Make sure no part of the cuff tube is
bent.