Cuff Selection And Connection - Omron HBP-1100 Instruction Manual

Professional blood pressure monitor
Hide thumbs Also See for HBP-1100:
Table of Contents

Advertisement

Cuff Selection and Connection

Warning
• Treat a cuff used on a patient with an infection as medical waste, or
sufficiently sterilize the cuff before reuse.
Caution
• Do not use the cuff if it is damaged or has holes.
• Use the appropriate cuff size to ensure correct measurements. If too
large a cuff is used, the measured blood pressure value tends to be
lower than the actual blood pressure. If too small a cuff is used, the
measured blood pressure value tends to be higher.
• Only an OMRON GS CUFF can be used with this device. The use of any
other cuff may result in incorrect measurement.
Note:
• It is important to use the correct sized cuff for a patient in order to get an
accurate reading.
Selecting the cuff
Measure the circumference of the patient's arm and select the cuff
size that is appropriate for the circumference.
Select the cuff that is suitable for the patient from the cuffs below.
GS CUFF XL*
GS CUFF L*
GS CUFF M
GS CUFF S*
GS CUFF SS*
*Available as an optional accessory.
Connecting the cuff
Connect the cuff tube to the NIBP connector
on the unit and turn clockwise to lock.
14
Cuff name
Arm circumference
(cm)
42 - 50
32 - 42
22 - 32
17 - 22
12 - 18
(inch)
17-20
13-17
9-13
7-9
5-7

Hide quick links: