(1)
X-Rite Calibration (Gradation Adjust)
1. Set up the Service Jig.
☞ See "Setting up the Service Jig." (D-2)
2. Call the Tech. Rep. mode to the screen.
☞ See "Calling the Tech. Rep. Mode to Screen." (D-5)
3. Click the X-Rite Calibration key in the Tech. Rep.
Mode screen.
C4581P004CA
4. Check the message that appears, and then click
the Start key in the External Panel Controller.
Test patterns are printed for the four colors.
(Cyan, Magenta, Yellow, Black)
NOTE
• It will take approximately 2 minutes for the test pat-
terns to be printed.
• During printing, do not open the front door.
If the front door is opened, restart the procedure
C4004P567CA
from step 1.
4004D279AA
5. After the test patterns are printed, check that the
message appears, and then click the Start key in
the External Panel Controller.
C4004P569CA
D-6