6.6

Adjusting the Tension of the Flip Up Footplate

The tension can be adjusted to increase or decrease
the rotation effort of the flip up footplates.
1.
Loosen the mounting screw on the front rigging angle hinge to
decrease the rotation effort.
DO NOT remove the footplate mounting screw.
2.
Tighten the front rigging angle hinge mounting screw to increase the
rotation effort.
3.
Repeat STEPS 1 and 2 for the other footplate.
6.7

Adjusting the Angle of the Cantilever Arm

Access to the adjustment screw is on the bottom of
the arm tube.
1.
Adjust the cantilever arm angle as necessary to accommodate the
user.
To adjust the arm angle down, turn the adjustment screw
clockwise.
To adjust the arm angle up, turn the adjustment screw
counterclockwise.
2.
Repeat STEP 1 to adjust the other cantilever arm (as necessary).
Part No 1143155
Center Mount
Front Rigging
Mounting Screw
FIGURE 10
Adjusting the Tension of the Flip Up Footplate
Back Cane
Arm Tube
FIGURE 11
Adjusting the Angle of the Cantilever Arm
71
6 ADJUSTMENT
Footplate
Arm Pad
Adjustment Screw
Invacare® Formula™ CG Seating