10. Toe In/Toe Out measurement is not within ±
11. Open the right camber clamp. Refer to Opening/Closing Camber Clamps on page 46.
12. Reposition the camber inserts to the highest degree of camber. Refer to Repositioning Camber Inserts (Adjusting Rear Wheel Camber). on
page 54.
13. Repeat STEP 3.
14. Close the left camber clamp. Refer to Opening/Closing Camber Clamps on page 46.
15. Rotate the right toe adjustment ring until the tab stops against the upper edge of the camber clamp.
16. Securely tighten set screw on right toe adjustment ring.
90° Angle
17. Repeat STEP 7.
Part No 1110545
1
/
inch - Repeat STEPS 1-7 until measurement is within ±
8
LOWER DEGREE OF CAMBER
Left Tab
Flat Edge of
Stopped
Camber Insert
against lower
edge of
"L" Square
camber clamp
Ground/Floor
Axle Tube will rotate up only
FIGURE 9 Adjusting The Axle Tube.
Left Toe
Adjustment
Ring
Right Tab Stopped against
upper edge of camber clamp
53
1
/
inch.
8
HIGHER DEGREE OF CAMBER
Axle Tube will rotate down only
Right Toe
Adjustment Ring
Invacare® A-4® Wheelchair