Save/Email/Recall (Setups, Screens, Data)
If the oscilloscope acquisition is stopped, data from the raw acquisition
record (which has more points than the measurement record) can be
written. Press the [Single] key to obtain maximum memory depth with
current settings. If enabled, serial decode data is saved.
When you want to save less than the maximum number of data points, a
1- of- N decimation is performed to produce an output whose length is less
than or equal to the requested length. For example, if there are 100k
points of data, and you specify a length of 2k, 1 of every 50 data points is
saved.
See Also

Length Control

The Length control is available when saving data to CSV, ASCII XY, or BIN
format files. It sets the number of data points that will be output to the
file. Only displayed data points are saved.
When Max Length is enabled, the maximum number of waveform data
points is saved.
The maximum number of data points depends on these things:
• Whether acquisitions are running. When stopped, data comes from the
• Whether the oscilloscope was stopped using [Stop] or [Single]. Running
• Whether only one channel of a pair is turned on. (Channels 1 and 2 are
• Whether reference waveforms are on. Displayed reference waveforms
• Whether digital channels are on. Displayed digital channels consume
• Whether segmented memory is on. Acquisition memory is divided by
312
"Binary Data (.bin) Format"
"CSV and ASCII XY files"
"Minimum and Maximum Values in CSV Files"
raw acquisition record. When running, data comes from the smaller
measurement record.
acquisitions split memory to provide fast waveform update rates. Single
acquisitions use full memory.
one pair, channels 3 and 4 are the other.) Acquisition memory is
divided among the channels in a pair.
consume acquisition memory.
acquisition memory.
the number of segments.