Base

The Base of a waveform is the mode (most common value) of the lower
part of the waveform, or if the mode is not well defined, the base is the
same as Minimum. The Y cursor shows the value being measured.

Overshoot

Overshoot is distortion that follows a major edge transition expressed as a
percentage of Amplitude. The X cursors show which edge is being
measured (edge closest to the trigger reference point).
Rising edge overshοot =
Falling edge overshοot = Base − D local Minimum
Agilent InfiniiVision 4000 X-Series Oscilloscopes User's Guide
local Maximum − D Top

Overshoot

local Maximum
local Minimum
Overshoot
Amplitude
Amplitude
Top

Base

14
Measurements
× 100