Setup And Hold Trigger - Agilent Technologies InfiniiVision 4000 X-Series User Manual

Hide thumbs Also See for InfiniiVision 4000 X-Series:
Table of Contents

Advertisement

Setup and Hold Trigger

The Setup and Hold trigger looks for setup and hold violations.
One oscilloscope channel probes the clock signal and another channel
probes the data signal.
To trigger on setup and hold violations:
1 Press the [Trigger] key.
2 In the Trigger Menu, press the Trigger softkey; then, turn the Entry knob
3 Press the Clock softkey; then, turn the Entry knob to select the input
4 Set the appropriate trigger level for the clock signal using the Trigger
5 Press the Rising Edge or Falling Edge softkey to specify the clock edge being
6 Press the Data softkey; then, turn the Entry knob to select the input
7 Set the appropriate trigger level for the data signal using the Trigger
8 Press the < Setup softkey, and turn the Entry knob to select the setup
Agilent InfiniiVision 4000 X-Series Oscilloscopes User's Guide
Data
Clock
(rising
edge)
to select Setup and Hold.
channel with the clock signal.
Level knob.
used.
channel with the data signal.
Level knob.
time.
Setup time
Hold time
10
Triggers
185

Advertisement

Table of Contents
loading

Table of Contents