Sexed Connectors; Visual Inspection - Agilent Technologies 346A Operating And Service Manual

Noise source
Hide thumbs Also See for 346A:
Table of Contents

Advertisement

Figure A-1
Dielectric
Support bead
Figure A-2
Precision 7mm Connector
Outer Conductor

Sexed Connectors

On sexed connectors, especially precision 3.5mm and SMA connectors,
pay special attention to the female center conductor contact fingers
(Figure A-2 and Figure A-3). These are very easily bent or broken, and
damage to them is not always easy to see. Any connector with damaged
contact fingers will not make good electrical contact and must be
replaced.
Precision 3.5mm connectors
FEMALE
Appendix A
Center Conductor
Outer
Conductor
Mating Plane
MALE
Outer Conductor
Mating Plane
Caring for Connectors

Visual Inspection

Collet
55