97. Heater SSRs swapping
If a faulty SSR is detected with the "SSR Fail" alarm this function can be used to swap the
SSR's to get the transmitter working.
"Swap SSR 1<–>2" will swap the two heater SSR's. Use this option if the SSR on heater
#1 fails and there is no probe connected to heater #2. The heater for probe #1 will have to
be plugged into the Heater #2 socket after this is changed.
"Swap Cal SSRs" will swap the two heater SSR's for the two cal gas solenoid SSR's. Use
this option if the transmitter is being used to control two probes but the calibration gas
solenoids are not being used.
Options:
Normal
*
Swap SSR 1<–>2
Swap Cal SSRs