Teledyne 9060 Operating Instructions Manual page 57

Zirconia oxygen analyzer
Table of Contents

Advertisement

9060-Zirconia Analyser
If 'AUTO CAL' is selected, the output channel is calibrated when 'Auto Cal' is initiated from
the keyboard (See Section 6.3).
If 'MAN CAL' is selected, it is necessary to trim both ends of the 4-20mA output range
using the 4mA and 20mA options in this menu item. Selecting 'MAN CAL' inhibits the
'Auto Cal' process of this channel.
Always do the 4mA trim first, and then the 20mA trim. After trimming both ends of the
scale, return the 'CALIBRATION OPTIONS' menu option back to 'MAN CAL' (not 'AUTO
CAL'), or the calibration factors will be over written by the next 'AUTO CAL'.
For more details on calibrating the output channels, see Section 6.3.
NOTE: The transmitter will only stay in either '4mA TRIM' or '20mA TRIM' modes for 30
minutes before it automatically returns to 'MAN CAL'.
14. Calibrate 4mA, Channel #1
This menu item only appears if 'Set 4mA Trim' is selected in Set-up 13.
Range: 0 to 25mA, Default is 4.00mA
For full details on the calibration of the 4-20mA output channels, see Section 6.3.
15. Calibrate 20mA, Channel #1
This menu item only appears if 'Set 20mA Trim' is selected in Set-up 13.
Range: 0 to 25mA, Default is 20.00mA
16. 4 to 20 mA Calibration Options, Channel #2
Select the calibration method for the 4-20mA output channel #1.
For more details, see Set-up 13 and Section 6.3.
Options:
1. Auto Calibration
*
2. Manual Calibration
3. Set 4mA Trim
11. Set 20mA Trim
17. Calibrate 4mA, Channel #2
This menu item only appears if 'Set 4mA Trim' is selected in Set-up 16.
Range: 0 to 25mA, Default is 4.00mA
For full details on the calibration of the 4-20mA output channels, see Section 6.3.
18. Calibrate 20mA, Channel #2
This menu item only appears if 'Set 20mA Trim' is selected in Set-up 16.
Range: 0 to 25mA, Default is 20.00mA
19. Enter Service Year
For a new 'DATE OF LAST SERVICE', enter the service 'YEAR'. This can represent the
last time the probe or sensor was serviced or the last time the boiler was serviced. It is
recommended that probes and sensors be refurbished every two years
20. Enter Service Month
Enter the current 'MONTH'.
57
57