Maintenance

5.3.6 UV Lamp (P/N ST20009)

Normal lamp replacement frequency is suggested at least every 12
months, however, if a lamp is "burned-out" (weak or no blue glow),
replace with UV Lamp (P/N ST20009), per directions in Figure 8.
To replace the UV lamp:
1. Turn off the UV lamp at the UV power supply. Remove
persulfate and acid reagent lines from containers, as well
as the sample inlet line and submerse all three (3) lines
into D. I. water to completely flush system.
CAUTION:
FAILURE TO DO SO MAY RESULT IN ACID BURNS,
CLOGGED REACTOR. FLUSH FOR 30 MINUTES.
CAUTION:
USE ONLY P/N ST20009 UV LAMP TO ASSURE
PROPER 185 MM AND 254 MM POWER SPECTRAL
DENSITY AND PROPER OXIDATION EFFICIENCY.
Teledyne Analytical Instruments

Figure 5-8: UV Lamp Replacement

Model 6750

56