Troubleshooting - Teledyne 6750 Operating Instructions Manual

Total organic carbon analyzer
Table of Contents

Advertisement

Maintenance
From this screen the operator may perform individual
testing of components related to functionality and
settings.

5.2 Troubleshooting

If you are experiencing trouble with your instrument, refer to Table
5-1: Troubleshooting.
Table 5-1: Troubleshooting
Analyzer will not turn on
Analyzer does not Calibrate, is
sluggish
Analyzer is excessively noisy
Display not functioning
Pump head not rotating
Pump head rotating but not
pumping fluid
Teledyne Analytical Instruments
PROBLEMS
Model 6750
SOLUTION
Check fuse & power connections.
Look for loose or damaged wiring.
Check for 12V DC from power supply.
Check power supply fuse.
Look for & correct any liquid or gas line
leaks.
Check for leaks at tubing connections
Check cable connections
Check power supply voltage
Suspect:
Malfunctioning Computer (P/N ST13039).
Master Control Panel Assembly (P/N
ST13042-1)
Refer to section "Pumps" 12.3.12
Verify that carrier loss did not cause the
pumps to be turned off.
Check for AC voltage at terminal strip on
inside of application panel.
Refer to section "PUMPS"
Change tubing
Verify sample line is in liquid.
44