3 Installation

Primary Input Power:

240V ac, 50/60 Hz power source. See Figure 3-4 for detailed connections.
DANGER: Power is applied to the instrument's circuitry as
long as the instrument is connected to the power
source. The standby function switches power on or
off to the displays and outputs only.
Tu rn cw to hold
ccw to
loo sen w ire.
In sert w ire
h ere.
Figure 3-4: Primary Input Power Connections

Fuse Installation:

type (slow blow) fuses. Fuses are not installed at the factory. Be sure to
install the proper fuse as part of installation (See Fuse Replacement in
chapter 5, maintenance.)
Analog Outputs:
the ANALOG OUTPUTS terminal block. There are two connectors per
output with the polarity noted. See Figure 3-5.
The outputs are:
0–1 V dc % of Range: Voltage rises linearly with increasing oxygen, from
0–1 V dc Range ID:
3-4: Part I
The universal power supply requires a 100-
The fuse holders accept 5 x 20 mm, 1.0 A, T
There are eight DC output signal connectors on
0 V at 0% to 1 V at 100%. (Full scale = 100%
programmed range.)
0.25 V = Low Range, 0.5 V = Medium Range,
0.75 V = High Range, 1 V = Air Cal Range.
Teledyne Analytical Instruments
Model 3010MB
H o t
G ro u n d
N eutral