Appendix - Teledyne 3010MB Operating Instructions Manual

Percent paramagnetic oxygen analyzer
Table of Contents

Advertisement

Percent Paramagnetic Oxygen Analyzer
A-1 Specifications
Sample System: Positive pressure service.
Response Time: 90% in 10 seconds or less.
Digital Interface: Full duplex RS-232 communications port.
Operating Temperature: 0-45 °C

Appendix

Packaging: Bulkhead mount.
Sensor: Paramagnetic Sensor.
Ranges: Three user definable ranges from 0-5 % to 0-
100 %.
Autoranging with range ID output.
Vacuum service (optional).
Auto Cal / Auto Zero. (Available with op-
tional, electrically operated valves.)
Alarms: One system-failure alarm contact to detect
power failure.
Two adjustable concentration threshold alarms
with fully programmable setpoints.
Diagnostics: Start-up or on-demand self-testing function.
Displays: 2 line by 20 alphanumeric, VFD screen, and
one 5 digit LED display.
Power: Universal power supply 100-240 V ac, at 50
or 60 Hz, 70 Watts max.
Accuracy: ±1% of full scale at constant temperature.
±5% of full scale over operating temperature
Teledyne Analytical Instruments
Appendix
A-1

Advertisement

Table of Contents
loading

Table of Contents