Features - Teledyne AX300 Operating/Service Instructions Manual

Portable oxygen analyzer
Hide thumbs Also See for AX300:
Table of Contents

Advertisement

Introduction
and a visual green light will flash, and the alarm buzzer will activate.
This alarm condition is also triggered when the oxygen concentration at
the point of analysis falls below 18% (Reference: ISO 7767 standard).
For safety reasons, the audible alarm cannot be silenced until the alarm
condition is resolved.
CAUTION:
DURING NORMAL USE, ALARM CONDITIONS SHOULD
NOT BE OVERRIDDEN BY ACTIONS SUCH AS HOLDING
THE ALARM SILENCE KEY DEPRESSED OR REMOVING
THE BATTERIES. THE CAUSE FOR THE ALARM CONDI-
TION SHOULD FIRST BE RESOLVED.
CAUTION:
WHILE HOLDING DOWN THE ALARM SILENCE KEY MAY
MOMENTARILY SILENCE THE ALARM BUZZER, AUDIBLE
ALARM WILL RETURN UPON DISENGAGING THE ALARM
SILENCE KEY. HOLDING THE KEY FOR LONGER DURA-
TIONS WILL INITIATE THE "STUCK KEY" ALARM, WHICH
CAN BE DEACTIVATED ONLY VIA REMOVING THE BAT-
TERIES FROM THE ANALYZER. ANALYZER SHOULD BE
RECALIBRATED UP ON REPOWERING THE ANALYZER.
NOTE:
THE ALARM SILENCE KEY DOES NOT PLAY ANY ROLE IN
THIS VERSION OF THE ANALYZER. IT IS DESIGNED IN
FOR ANY POSSIBLE FUTURE PRODUCT MODIFICATION.

1.1 Features

The following features are standard on the AX300:
Large easy to read 3½-digit LCD (see options)
Automatic LCD back lighting upon key press
Microprocessor controlled
Sensor fail/disconnect alarm indicator (audible and visual)
Non-adjustable concentration alarm below 18% O
2000 operating hours from 3 AA alkaline batteries
Battery status indicator
Stand for upright tabletop deployment
Optional hardware for pole clamping and V block support
Rugged high impact ABS plastic construction
Splash resistant case
Long life class R17MED sensor
Teledyne Analytical Instruments
AX300