Service Instructions

3.4 Sensor Maintenance

DO check the sensor for damage or leaks before use.
DON'T immerse the R17MED sensor in liquid.
DON'T autoclave the R17MED sensor.
DON'T open or try to repair the sensor.
Before every use, the sensor, cable and connections should be checked.
Check the sensor for leaks and condensation. Check the cable for splitting or
cracked insulation. Make sure the connections are tight and dry.
In the event that the sensor has been damaged, consult the Material Safety
Data Sheet in the Appendix for handling guidelines.

3.5 Calibration

Incorrect readings can often be traced to improper calibration. The AX300
should be calibrated before each use and every 8 hours to maintain accuracy. It
must be calibrated whenever new sensor or batteries are installed. Calibration
using methods other than described in section 2.1.4 will lead to improper opera-
tion.
Whenever new batteries are installed or removed for any reason, the oxygen
analyzer defaults to the calibration mode with the LCD display flashing 00.0.
Only the CAL and ON/OFF (I/O) keys are functional at this point.

3.6 Factors Affecting Sampling Accuracy

Refer to Section 2.3 of this manual.

3.7 Troubleshooting

The AX300 oxygen analyzer provides a variety of built-in safety features that
prevent its use when a fault is detected. When a unit displays the message √ SENSOR
and sounds the audible and visual alarm continuously, it is an indication of a faulty
connection between the sensor and the unit or an expired or faulty sensor. It also indi-
cates when oxygen concentrations are below 18%.
For assistance with troubleshooting, see Table 3-1.
Teledyne Analytical Instruments
AX300
22