AC Current

Press the ACI key to show the ACI menu:
ACI
Auto
ACI Configuration (Resolution, RMS Filtering and DC Coupled)
Press the
Resl
Filt
DCcp
ACI Resolution
Press the
This menu offers the following resolution choices:
5
6
To revert to the
3-22
This function provides measurements, in decade ranges from 200 A to
20A. (NOTE: the 20A range is only available on the front terminals. The
front current terminal is not fused.)
To cancel autoranging and enter a manually selected range, press any range
selection or the Auto softkey. The Multimeter enters the selected range or
reverts to the auto-selected range.
key to see the
Config
Provides selection of the resolution for subsequent readings.
Provides selection of various filters for the RMS converter, allowing
measurements to be made down to the chosen filter frequency. One of four
filters is always in circuit. The 40Hz filter is the default selection at power
on.
The Filter annunciator on the left-hand display indicates that the filter is
active and indicates its frequency.
Controls the measurement-coupling configuration.
The DC coupled annunciator in the left-hand display indicates active state.
key to see the
Resl
5-1/2-digits resolution
6-1/2-digits resolution
ACI CONFIG
ACI Menu
menu:
ACI CONFIG
ACI CONFIG Menu
menu:
ACI RESL
ACI RESL
Menu
menu, press the
key.
Config
adj101f.eps
adj102f.eps
adj103f.eps