Division - Fluke 8508A User Manual

Reference multimeter
Hide thumbs Also See for 8508A:
Table of Contents

Advertisement

Recall Subtraction Constant
C?
(Nr3) Recalls the defined value of C.

Division

Each signal value is divided by a user-defined factor "Z".
Enable Division
Selects the division operation to be performed on the measurement.
Power On and Reset Conditions
OFF
Set Division Constant
This command allows the definition of the factor Z.
Nrf
A numeric value required for use in the DIV_Z processing. The decimal data resolution is
8.5 digits; numbers exceeding this resolution will be rounded to 8.5 digits. Divide by
zero will set bit 5 (MOF) of the Measurement Event Status Byte. This value is saved in
non volatile memory.
Example:
Z -56.999
LAST_RDG
Places the most recent reading into the numeric memory.
The mathematical processing capability is limited in the range of numbers which it can
successfully handle. The maximum resolution of the mantissa is 8.5 digits, and the
exponent is limited to ±15. Calculations which result in values outside this range will
produce an error indicated by the invalid response when accessed by a query command.
C?
Recall Subtraction Constant
phs
DIV _Z
Enable Division
Z
phs
Set Division Constant
sets Z to -56.999.
Remote Operations Using the IEEE 488 Interface
Syntax Diagrams of Math Functions
ON
OFF
Nrf
LA ST_RDG
4
adj234f.eps
adj235f.eps
adj236f.eps
4-29