RESL5 / RESL6 / RESL7 / RESL8
Sets the resolution of the measurement to 5.5, 6.5, 7.5, or 8.5, digits.
FAST_ON / FAST_OFF
FAST_ON
TWO_WR
Selects 2-wire Ohms (Use Hi and Lo terminals). (For backwards compatibility TWR can
also be used.)
FOUR_WR
Selects 4-wire Ohms. (For backwards compatibility FWR can also be used.)
LOI_ON / LOI_OFF
Selects low current mode.
Power On and Reset Conditions
20 k , FILT_OFF, RESL7, FAST_ON, TWO_WR, LOI_OFF

Resistance: High Voltage OHMS

The following commands are used to select the High Voltage
configuration.
Nrf
A numeric value that selects the range for the expected signal. Any valid numeric value
cancels autorange.
FILT_ON / FILT_OFF
FILT_ON
RESL5 / RESL6 / RESL7 / RESL8
Sets the resolution of the measurement to 5.5, 6.5, 7.5, or 8.5, digits.
FAST_ON / FAST_OFF
FAST_ON
4-14
Reduces the A-D integration time, for faster conversions.
HIV_OHMS
phs
High Voltage OHMS
inserts a hardware analog filter into the measurement signal path.
Reduces the A-D integration time, for faster conversions.
OHMS
,
Nrf
FILT_ON
FILT_OFF
RESL5
RESL6
RESL7
RESL8
TWO_WR
FOUR_WR
FAST_ON
FAST_OFF
function and set its
adj213f.eps