Cal Due Date Entry And Calibration Mode Disable; Acv Spot Frequency Calibration - Fluke 8508A User Manual

Reference multimeter
Hide thumbs Also See for 8508A:
Table of Contents

Advertisement

8. Press the
CONFIG
9. Press the
CONFIG
10. Press the
PLUS
11. Press the
CONFIG
12. Press the
CONFIG
13. Press the
PLUS
14. Press the
CONFIG
15. Press the
CONFIG
16. Press the
DCV

Cal Due Date Entry and Calibration Mode Disable

On completion of all calibration operations it is recommended that the user updates the
stored
due date with the date for next recalibration. For consistency, this date should
CAL
be the same as the calibration due date shown on any calibration stickers the user may
also attach to the instrument. It is also recommended that a calibration seal or integrity
sticker is applied to prevent unauthorized access to the rear panel calibration enable
switch.
Set Cal Due Date
1. Press the
CAL
2. From the
CAL DUE ?
3. Use the keypad to key in the calibration due date. Up to eight characters may be
entered in any format. Any character available from the keypad may be used, but it is
recommended that one of the common date formats such as dd.mm.yy or mm.dd.yy
is used for ease of interpretation.
4. Press Enter to store the
Disable cal mode
1. Set the rear panel Calibration Enable switch to
2. Apply a security seal to cover the rear panel Calibration Enable switch to prevent
unauthorized access.

ACV Spot Frequency Calibration

The ACV Spot frequency mode calibration is not normally performed as part of routine
calibration. ACV Spot frequency calibration is only performed if the user intends to make
use of spot frequency mode. In spot frequency mode the accuracy of AC Voltage
measurements can be improved by calibrating the 8508A at specific frequencies of
interest, reducing frequency response flatness errors at these points. The following
procedure provides guidance on spot frequency calibration for users wishing to make use
of Spot frequency mode.
key, select
and de-select
4w
key and select
2w
key and select
. From the
Tru
key and select
and
Filt
key and de-select
LoI
key and select
. From the
HiV
key and select
and
Filt
key and select
4w
key and remove the shorting devices from the input terminals.
key and select Quit.
menu select Enter.
due date.
CAL

ACV Spot Frequency Calibration

. Repeat steps 3 and 4.
LoI
. Repeat steps 3 and 4.
menu the 2
Tru
. Repeat steps 3 and 4.
LoI
. Repeat steps 3 and 4.
menu the 20 M range.
HiV
. Repeat steps 3 and 4.
2w
. Repeat steps 3 and 4.
.
DISABLE
Calibration and Verification
range.
6
6-29