Table 6-9. AC Voltage Verification Points and Sequence
8508A Range
Voltage
200 mV
100 mV
100 mV
100 mV
100 mV
100 mV
100 mV
100 mV
100 mV
2 V
20V
10 V
10 V
10 V
10 V
10 V
10 V
10 V
10 V
10 V
10 V
19 V
Frequency
20 Hz
55 Hz
1 kHz
3 kHz
10 kHz
30 kHz
60 kHz
100 kHz
1 V
20 Hz
1 V
55 Hz
1 V
1 kHz
1 V
3 kHz
1 V
10 kHz
1 V
30 kHz
1 V
60 kHz
1 V
100 kHz
1 V
500 kHz
1 V
1 MHz
1 V
1 kHz
20 Hz
55 Hz
1 kHz
3 kHz
10kHz
30 kHz
60 kHz
100 kHz
500 kHz
1 MHz
1 kHz
Calibration and Verification
Performance Verification
8508A Filter
Tolerance
10 Hz
40 Hz
100 Hz
100 Hz
100 Hz
100 Hz
100 Hz
100 Hz
10 Hz
0.000144 V
40 Hz
0.000119 V
100 Hz
0.000099 V
100 Hz
0.000119 V
100 Hz
0.000119 V
100 Hz
0.000295 V
100 Hz
0.000795 V
100 Hz
0.000795 V
100 Hz
0.034000 V
100 Hz
0.034000 V
100 Hz
10 Hz
40 Hz
100 Hz
100 Hz
100 Hz
100 Hz
100 Hz
100 Hz
100 Hz
100 Hz
100 Hz
1
0.0195 mV
0.0175 mV
0.0149 mV
0.0175 mV
0.0175 mV
0.0445 mV
0.0995 mV
0.0995 mV
0.00032 V
0.00144 V
0.00119 V
0.00099 V
0.00119 V
0.00119 V
0.00295 V
0.00795 V
0.00795 V
0.34000 V
0.34000 V
0.00167 V
6
6-35