Refer to table 3-1 for the ADC integration times applicable to the various resolution and
Fast On/Off settings.
To revert to the
TRU - Movement between Menus

Measuring Resistance

2-Wire Measurements

For many applications the simple 2-wire arrangement will be adequate. See Figure 3-5.
However, the value displayed will include the resistance of the connecting leads.
Use a shielded twisted-pair cable, preferably of PTFE insulation, to reduce induced
voltages, induced charge and shunt leakage resistance, particularly where Rx is high.
2-wire resistance measurement is not available when in
to use in the 2
resistance contributions may limit full-scale readout. 2-wire measurements above 1.5
should be made using higher ranges.
3-16
key to see the
Resl
5-1/2-digits resolution
6-1/2-digits resolution
7-1/2-digits resolution
8
1/2-digits resolution
-
TRU CONFIG
TRU
range. In the latter case zero compensation for lead and internal
menu:
TRU RESL
TRU RESL
Menu
menu, press the
Config
Movement Between
Menu
key.
mode and is not well suited
Tru
adj083f.eps
adj086f.eps