16.2.
16.2.
16.2.
16.2.
There are two data format types set at F-Function f34 and f35.
The initial data format of f34 is shown below.
S
T
,
Header 1 Header 2
S
T
U
S
O
L
H
D
Header 1 f34, 35 : 9 (Status)
ST
Stable weight data.
US
Unstable weight data.
OL
Overload (Out of range).
HD
Held weight data (The response of the "RW and "RW,n" command).
f34, 35 : e (weight type)
Header 2
GS
G
or
NT
N
or
TR
T
or
f34, 35 : a, b, C, d (weight data)
Data
The first of the data bits is the polarity, "+" or "-".
When data is zero, the polarity is "+".
8 digits including polarity and decimal point (dot or comma).
In case of "out of range", the data are replaced by spaces except for the decimal
point.
f34, 35 : f (unit)
Unit
In case of Cf07 0, the unit length is 2 digits.
In case of Cf07 1, the unit length is 3 digits. Depending on circumstances, an A&D
printer may not work correctly.
AD-4405 Weighing Indicator

Data Format

Data Format
Data Format
Data Format
G
S
,
+
0
0
0
0
0
8 digit data
G
S
G
N
T
N
T
T
R
P
P
T
T
Cf06 = 0 , 1 Cf06 = 2
Gross data.
Net data
Tare data.
.
0
k
g
Terminators
Unit
g
g
k
k
g
g
t
t
P
C
P
C
Cf07 = 0 Cf07 = 1
2 figures 3 figures
Page 57
Space code(20h)
Carriage return(0Dh)
Line feed(0Ah)
Decimal point :
Dot(2Eh) or
Comma(2Ch)
l
b
l
b
o
z
o
z
c
y
c
y
t
l
t
l
16.2. Data Format