Step 3 After displaying Cap for 2 seconds, the first range or the weighing capacity of
single range will be displayed. When dual range is used, Cap1 is displayed for 2
seconds. Triangle
and press the
the
key, regardless of what is displayed, the indicator will proceed to the next
ESC
step without changing the parameters. The next step is Zero Calibration in single
range or the second range resolution in dual range.
Specifying the Second Range Resolution
Specifying
Specifying
Specifying
the Second Range Resolution
the Second Range Resolution
the Second Range Resolution
Step 4 After displaying range2 for 2 seconds, the resolution with decimal point and triangle
2 will be displayed. Specify the second range resolution in the same way as the
first range. The decimal point cannot be moved. Specify the second range resolution
greater than the first range. Press the
proceed to the next step. When pressing the
displayed, the indicator will proceed to the next step without changing the parameter.
Specifying
Specifying the Second Range Capacity
Specifying
Specifying
the Second Range Capacity
the Second Range Capacity
the Second Range Capacity
Step 5 After displaying Cap2 for 2 seconds, the capacity with unit and decimal point is
displayed. Specify the capacity in the same way as the first range. The capacity
should be greater than the first range. Press the
and proceed to Zero Calibration.
5.2.2.

To Get Stabilized Data

Step 6 Maintain the following conditions to calibrate the scale (indicator) correctly.
Maintain a constant temperature, stable power and stable input voltage from the
load cell.
Avoid direct sunshine or the near the outlet of an air conditioner.
Do not install the scale (indicator) where there is a strong magnetic field.
Step 7 Turn the display on and leave it for several minutes.
5.2. Calibration Procedure
1 will be displayed. Specify parameter with the
key to store it and proceed to the next step. When pressing
ENTER
key to store the parameters and
ENTER
ESC
ENTER
Page 14
0
key, regardless of what is
key to store the parameters