5.6. Calibration Error Code List - A&D AD-4405 Instruction Manual

Weighing indicator
Table of Contents

Advertisement

5.6. Calibration Error Cod
5.6.
Calibration Error Code List
5.6.
5.6.
Calibration Error Cod
Calibration Error Cod
Exiting from a Calibration error
key..... The key to return to the point where an error occurred. Retry the operation.
ESC
key while pressing the
ESC
No parameters are changed, CanCel is displayed and the calibration mode
is finished. Press the
Error Code List
If an error has occurred during the calibration mode, the following code is displayed.
Error code
In multi-interval scale. The last division is set to maximum ( d-50 ).
err 0
Therefore the next division can not be entered.
Resolution exceeds 40000. (Resolution = maximum capacity/ minimum
err 1
division)
Reduce the maximum capacity or increase the minimum division.
Load cell output is too large or too small at zero calibration. Check the weighing
err 2
unit and load cell. Refer to "3.1.4.Verifying Load Cell Output and Input
err 3
Sensitivity".
Measuring calibration mass, the value exceeded maximum capacity. Reduce
err 4
the calibration mass.
err 5
The selected calibration mass is smaller than the minimum division.
The new input sensitivity is less than 0.2 µV/division. Increase the input
err 6
sensitivity. Refer to "3.1.4.Verifying Load Cell Output and Input Sensitivity".
Placing a mass on the weighing unit, the load cell output becomes a negative
err 7
value. Check the load cell cable connections and the direction of load cell
mounting.
The load cell output exceeds the input range before the maximum capacity.
Adjust zero balance referring to "3.1.4.Verifying Load Cell Output and Input
err 8
Sensitivity". Replace with a load cell designed for a smaller output. Reduce
maximum capacity.
The weight value is out of the input range at zero calibration or span
err 9
calibration. Check the weighing unit and cables.
err 12
The first weighing range is larger than second weighing range.
An incorrect mass was selected at the digital linearization function.
err 13
Select a mass of the following relation. Lnr 1 < Lnr 2 < Lnr 3.
g err
An unacceptable value was selected in the gravity acceleration function.
AD-4405 Weighing Indicator
e List
e List
e List
key.
ON/OFF
key to turn the display off.
ON/OFF
Description
Page 21

5.6. Calibration Error Code List