Replacing The Device - Siemens SITRANS TDL Compact Operating Instructions

In situ laser gas analyzers
Hide thumbs Also See for SITRANS TDL:
Table of Contents

Advertisement

Hot parts in the device
Temperatures that can burn unprotected skin may be present for some time after the device
has been switched off.
• Observe the waiting time specified in Technical specifications (Page 52) or on the device
Hazardous voltage at open device
Risk of electric shock when the enclosure is opened or enclosure parts are removed.
• Before you open the enclosure or remove enclosure parts, de-energize the device.
• If maintenance measures in an energized state are necessary, observe the particular
8.4

Replacing the device

Corrosive substances
Risk of chemical burns when replacing the sensor.
The sensor in the device contains corrosive substances that result in burns on unprotected
skin.
• Make sure that the sensor enclosure is not damaged when replacing the sensor.
• If contact with the corrosive substances occurs, rinse the affected skin immediately with
SITRANS TDL
Compact Operating Instructions, 05\2023, A5E51659183-AA – revised 03/2024
CAUTION
before starting with maintenance work.
CAUTION
precautionary measures. Have maintenance work carried out by qualified personnel.
CAUTION
large amount of water to dilute substance.
Service and maintenance
8.4 Replacing the device
49

Advertisement

Table of Contents
loading