Siemens SITRANS TDL Operating Instructions Manual

In situ laser gas analyzers
Hide thumbs Also See for SITRANS TDL:
Table of Contents

Advertisement

Quick Links

Continuous gas analysis
In situ Laser Gas Analyzers
SITRANS TDL
Operating Instructions
7MB6431 (SITRANS TDL)
05/2023
A5E51429088_AA
Getting started
Introduction
Safety notes
Description
Installing/mounting
Connecting
Commissioning
Operating
Parameter assignment
Service and maintenance
Diagnostics and
troubleshooting
Technical specifications
Dimension drawings
Product documentation
and support
Technical reference
Communication
Remote operation
1
2
3
4
5
6
7
8
9
10
11
12
13
A
B
C
D

Advertisement

Table of Contents
loading

Summary of Contents for Siemens SITRANS TDL

  • Page 1 Description In situ Laser Gas Analyzers SITRANS TDL Installing/mounting Connecting Operating Instructions Commissioning Operating Parameter assignment Service and maintenance Diagnostics and troubleshooting Technical specifications Dimension drawings Product documentation and support 7MB6431 (SITRANS TDL) Technical reference Communication 05/2023 Remote operation A5E51429088_AA...
  • Page 2 Note the following: WARNING Siemens products may only be used for the applications described in the catalog and in the relevant technical documentation. If products and components from other manufacturers are used, these must be recommended or approved by Siemens. Proper transport, storage, installation, assembly, commissioning, operation and maintenance are required to ensure that the products operate safely and without any problems.
  • Page 3: Table Of Contents

    3.1.5 Improper device modifications.................... 17 Requirements for special applications................. 17 Description............................19 Overview SITRANS TDL....................... 19 System configuration ......................20 SITRANS TDL mechanical assembly ..................21 Applications........................22 Approvals .......................... 23 Installing/mounting ..........................25 Basic safety notes....................... 25 5.1.1 Installation location requirements ..................26 5.1.2...
  • Page 4 Table of contents 5.2.8.3 Installing process flanges ....................36 5.2.8.4 Flange installation ......................37 5.2.8.5 Alignment of SITRANS TDL ....................39 5.2.8.6 Alignment kit ........................40 5.2.8.7 Alignment procedure ......................40 Sensor mounting ....................... 44 5.3.1 Mounting conditions ......................44 5.3.2...
  • Page 5 Limit 2 (2.4.2) to Limit 10 (2.4.10)................... 119 9.2.5 Noise reduction (2.5) ....................... 119 9.2.5.1 Damping value T90 (2.5.1) ....................119 9.2.6 Inputs and outputs (2.6) ....................119 9.2.6.1 Current input (2.6.1)......................119 9.2.6.2 Current output (2.6.2) ..................... 120 SITRANS TDL Operating Instructions, 05/2023, A5E51429088_AA...
  • Page 6 Enable user PIN (5.5) ....................... 132 9.5.6 Auto logout (5.7) ......................132 9.5.7 Logout (5.8) ........................132 Service and maintenance ........................133 10.1 Basic safety notes......................133 10.1.1 Maintenance........................133 10.1.2 Dust layers above 5 mm ....................133 SITRANS TDL Operating Instructions, 05/2023, A5E51429088_AA...
  • Page 7 Principle of operation ...................... 153 Calibration and calibration certificate ................153 Diagnostic LEDs ....................... 154 Communication ..........................155 Modbus RTU ........................155 C.1.1 Introduction........................155 PROFIBUS......................... 155 C.2.1 Introduction........................155 C.2.2 PROFIBUS communication....................155 SITRANS TDL Operating Instructions, 05/2023, A5E51429088_AA...
  • Page 8 Modbus TCP ........................157 SIMATIC PDM ........................158 D.3.1 Overview SIMATIC PDM ....................158 D.3.2 Updating the Electronic Device Description (EDD) or Field Device Integration (FDI).... 159 D.3.3 Check SIMATIC PDM version ..................... 159 Index ..............................161 SITRANS TDL Operating Instructions, 05/2023, A5E51429088_AA...
  • Page 9: Getting Started

    Read the entire document in order to achieve the optimum performance of the device. Procedure 1. Install/mount the device. Installating/mounting (Page 25) 2. Connect the device. Connecting (Page 49) 3. Power up the device. Device startup (Page 73) 4. Commission the device. Commissioning (Page 71) Startup is complete. SITRANS TDL Operating Instructions, 05/2023, A5E51429088_AA...
  • Page 10 Getting started 1.1 Getting started SITRANS TDL Operating Instructions, 05/2023, A5E51429088_AA...
  • Page 11: Introduction

    Compatible version of device integration pack‐ ual ed‐ ition 04/202 First edition HMI-FW: 5.06.14-28 or lat‐ SIMATIC PDM: 8.2 SP1 or EDD: 1.00.00 or later later HMI-MS: 0.00.00-28 or lat‐ Device revision A01 or lat‐ SITRANS TDL Operating Instructions, 05/2023, A5E51429088_AA...
  • Page 12: Items Supplied

    Introduction 2.2 Items supplied Items supplied • SITRANS TDL gas analyzer               • Siemens Process Analytics documentation disk containing certificates, and manuals. (All device manuals and certificates can be found by scanning the QR code on the de‐...
  • Page 13: Security Information

    Siemens’ products and solutions undergo continuous development to make them more secure. Siemens strongly recommends that product updates are applied as soon as they are available and that the latest product versions are used. Use of product versions that are no longer supported, and failure to apply the latest updates may increase customer’s exposure...
  • Page 14: Notes On Warranty

    The content reflects the technical status at the time of publishing. Siemens reserves the right to make technical changes in the course of further development.
  • Page 15: Safety Notes

    Observe the test certification, provisions and laws applicable in your country during connection, assembly and operation. These include, for example: • National Electrical Code (NEC - NFPA 70) (USA) • Canadian Electrical Code (CEC Part I) (Canada) SITRANS TDL Operating Instructions, 05/2023, A5E51429088_AA...
  • Page 16: Conformity With European Directives

    The product described in this document is in conformity with the relevant harmonization legislation, and its amendments, of the United Kingdom. The applicable regulations can be found in the UKCA declaration of conformity of the specific device. SITRANS TDL Operating Instructions, 05/2023, A5E51429088_AA...
  • Page 17: Improper Device Modifications

    Note Operation under special ambient conditions We highly recommend that you contact your Siemens representative or our application department before you operate the device under special ambient conditions as can be encountered in nuclear power plants or when the device is used for research and development purposes.
  • Page 18 • If no settings need be carried out, lock the device. SITRANS TDL is classified as a Class 1 laser product. The emitted laser radiation is invisible (near infrared) and not hazardous to an unprotected eye. SITRANS TDL has warning labels placed at positions specified in the EN 60825-1:2014-11 norm.
  • Page 19: Description

    Description Overview SITRANS TDL The laser based SITRANS TDL analyzer is a versatile tunable diode laser (TDL) gas analyzer for industrial process applications. SITRANS TDL is normally mounted directly on the process using a pair of process flanges. SITRANS TDL...
  • Page 20: System Configuration

    4.2 System configuration System configuration The SITRANS TDL analyzer consists of a pair of cross-duct sensors - an emitter unit and a receiver unit. The emitter unit releases laser radiation directly through the process containing the constituents of interest. The receiver unit collects the laser radiation on the other side of the process duct.
  • Page 21: Sitrans Tdl Mechanical Assembly

    Description 4.3 SITRANS TDL mechanical assembly SITRANS TDL mechanical assembly SITRANS TDL device assembly Receiver unit     Emitter unit ① ⑧ Unit housing     Purging tubes ② ⑨ Cable glands     Process flange ③ ⑩ Tube    ...
  • Page 22: Applications

    Modbus cable gland     Applications SITRANS TDL is designed for use in many industrial applications such as: • Power industry – monitoring of the combustion process • Wood board production – operator safety, emission monitoring • Chemical industry • Cement industry •...
  • Page 23: Approvals

    Description 4.5 Approvals Approvals In all cases, check the device nameplate and confirm the approval rating. Note Full list of approvals For more information, see Technical specifications (Page 141). SITRANS TDL Operating Instructions, 05/2023, A5E51429088_AA...
  • Page 24 Description 4.5 Approvals SITRANS TDL Operating Instructions, 05/2023, A5E51429088_AA...
  • Page 25: Installing/Mounting

    Note Material compatibility Siemens can provide you with support concerning selection of parts wetted by process media. However, you are responsible for the selection of parts. Siemens accepts no liability for faults or failures resulting from incompatible materials. WARNING Unsuitable connecting parts Risk of injury or poisoning.
  • Page 26: Installation Location Requirements

    The device may overheat if there is an insufficient supply of air. • Install the device so that there is sufficient air supply in the room. • Observe the maximum permissible ambient temperature. Refer to the information in the section Technical specifications (Page 141). SITRANS TDL Operating Instructions, 05/2023, A5E51429088_AA...
  • Page 27: Proper Mounting

    • Before installing ensure there is no visible damage to the device. • Make sure that process connectors are clean, and suitable gaskets and glands are used. • Mount the device using suitable tools. Refer to the information in Technical specifications (Page 141). SITRANS TDL Operating Instructions, 05/2023, A5E51429088_AA...
  • Page 28: Installation Instructions

    [2:3:3] according to ISO 8573-1:2010, recommended Class 2 or 3. Bad quality of purging gas may affect the measurement or even damage the analyzer. ① Housing purge ② Tube purge ③ Process purge SITRANS TDL Operating Instructions, 05/2023, A5E51429088_AA...
  • Page 29: Housing Purging (Optional)

    5.2.2 Housing purging (optional) SITRANS TDL device allows purging of the interior of the electronics housing by using inlets/ outlets found on the bottom side of each unit. 1. Prior to the inserting the tubes into the fittings, ensure openings are cleared from obstruction that can disrupt purge gas flow.
  • Page 30: Tube And Process Purging

    6 mm (0.25 in) double stainless steel 316 compression fitting 5.2.3 Tube and process purging Note Purging the emitter and receiver The process of purging is the same on the emitter side and receiver side. SITRANS TDL Operating Instructions, 05/2023, A5E51429088_AA...
  • Page 31 Process purging should be supplied by another gas source or connected in series with tube and housing purging via check valve to prevent any process gas access into the device tubes or housings. Arrows indicate purge gas flow direction. SITRANS TDL Operating Instructions, 05/2023, A5E51429088_AA...
  • Page 32 ① Purge gas source two ② To another tube or housing (or exhaust) ③ Purge gas source one Option two: Purging SITRANS TDL unit and process from the same gas source ① Check valve ② Purge gas source Note Process for tube and housing purging When process and tube and housing purging are connected in series, it is necessary to complete the purging gas flow rates.
  • Page 33: Purging Tubes

    See flange installation (Page 37) for installing graphite gasket on the purging tube. ① Process flange ② Graphite gasket ③ Purging tube ④ Graphite gasket ⑤ Process interface SITRANS TDL Operating Instructions, 05/2023, A5E51429088_AA...
  • Page 34: Emitter Unit

    Maximum angle of central unit lid opening   5.2.6 Receiver unit Cautiously dismount housing lid by unscrewing each of the four bolts. The receiver unit lid is only attached by the bolts. SITRANS TDL Operating Instructions, 05/2023, A5E51429088_AA...
  • Page 35: Emitter Unit - Electrical Terminals

    5.2.8.1 Safety information SITRANS TDL is classified as a Class 1 laser product. The emitted laser radiation is invisible (near infrared) and not hazardous to an unprotected eye. SITRANS TDL has warning labels placed at positions specified in the EN 60825-1:2014-11 norm.
  • Page 36: Change Of Laser Class

    +-1° to enable optimum (following) alignment of sensors to the optical axis. The emitter unit and the receiver unit need to be aligned with respect to the axis of symmetry of the flange. SITRANS TDL Operating Instructions, 05/2023, A5E51429088_AA...
  • Page 37: Flange Installation

    Dimension B ④ Max. deviation ± 1°     During the installation of the SITRANS TDL, check that the light from the laser inside the emitter reaches the photodiode inside the receiver by using the alignment procedure. (Page 40) 5.2.8.4 Flange installation...
  • Page 38 M16 Screws mounting     Mount flange on the proc‐ ess flange with another graphite gasket between flange and purging tube. ETC- mounting flange on the process flange SITRANS TDL Operating Instructions, 05/2023, A5E51429088_AA...
  • Page 39: Alignment Of Sitrans Tdl

    ETC clamp ring 5.2.8.5 Alignment of SITRANS TDL The emitter and receiver units must be aligned using two adjustment screws. The adjustment screws align the emitter and receiver units, whereby the laser beam from the emitter unit reaches the receiver unit, and the receiver unit that faces the emitter unit detects the laser beam.
  • Page 40: Alignment Kit

    Flange purging is recommended for every installation and should be operational before installation of the flange on the process. This ensures that the windows will remain clean during the installation of the flanges. SITRANS TDL Operating Instructions, 05/2023, A5E51429088_AA...
  • Page 41 4. Position the aiming piece base on the flange and en‐ sure the screw on the side of the aiming tool is in line with the purging inlet (on the side of the flange). Aiming piece positioning SITRANS TDL Operating Instructions, 05/2023, A5E51429088_AA...
  • Page 42 Light spot visibility 8. By sliding the coarse align‐ ing unit, the bright spot may be focused on the screen. Find a position at which the spot becomes the smallest and sharpest. Finding the spot SITRANS TDL Operating Instructions, 05/2023, A5E51429088_AA...
  • Page 43 (both emit‐ ter or receiver unit). Finding the optimal aligning point. The optimal aligning point depends on the device variant. ① Incorrect ② Correct SITRANS TDL Operating Instructions, 05/2023, A5E51429088_AA...
  • Page 44: Sensor Mounting

    If these conditiions cannot be fulfilled, a heat protection must be installed. Depending on local conditions, solutions may range from installation of a sun shield to operating SITRANS TDL in a cabinet with controlled environment.
  • Page 45: Optical Path Length Determination

    Correct optical path length is necessary for proper device performance. User must enter the actual value of the measuring path length in meters via SITRANS TDL HMI or via the web server. See parameter Path length (2.3) (Page 117). The measuring path length is defined as the distance between the ends of purging tubes.
  • Page 46: Sensor Mounting

    The connection can be tightened now. Receiver-flange connection 4. To tighten the con‐ nection, place the clamp on the rim of the tube. Placing the clamp SITRANS TDL Operating Instructions, 05/2023, A5E51429088_AA...
  • Page 47: Disassembly

    • If the device contains hazardous media, it must be emptied prior to disassembly. Make sure that no environmentally hazardous media are released. • Secure the remaining connections so that no damage can result if the process is started unintentionally. SITRANS TDL Operating Instructions, 05/2023, A5E51429088_AA...
  • Page 48 Installing/mounting 5.4 Disassembly SITRANS TDL Operating Instructions, 05/2023, A5E51429088_AA...
  • Page 49: Connecting

    Improper power supply • Connect the device in accordance with the specified power supply and signal circuits. The relevant specifications can be found in the certificates, in Technical specifications (Page 141) or on the nameplate. SITRANS TDL Operating Instructions, 05/2023, A5E51429088_AA...
  • Page 50 Electromagnetic compatibility (EMC) You can use this device in industrial environments, households and small businesses. Metal enclosures ensure improved electromagnetic protection from high frequency radiation. This protection can be increased by grounding the enclosure. SITRANS TDL Operating Instructions, 05/2023, A5E51429088_AA...
  • Page 51: Improvement Of Interference Immunity

    • Keep device and cables at a distance from strong electromagnetic fields. • Take account of the conditions for communication specified in the Technical specifications (Page 141). • Use shielded cables to guarantee the full specification according to Modbus/Profibus DP. SITRANS TDL Operating Instructions, 05/2023, A5E51429088_AA...
  • Page 52: Connecting Sitrans Tdl

    6.2.1 Electrical connections The SITRANS TDL gas analyzer consists of a cross-duct sensors - a emitter unit and a receiver unit. The emitter unit emits laser radiation directly through the process containing the constituents of interest. The receiver unit collects the laser radiation on the other side of the process duct. The receiver unit is connected to the emitter unit by means of a loop cable.
  • Page 53: Available Interfaces

    Connecting 6.2 Connecting SITRANS TDL 6.2.2 Available interfaces Available interfaces • 4 x analog output 4 – 20 mA (isolated, easy user selection via DIP switch between active/ passive mode) • 4 x analog input 4 – 20 mA (easy user selection via DIP switch between active/passive mode) •...
  • Page 54: Emitter Unit - Electrical Terminals

    Connecting 6.2 Connecting SITRANS TDL Row Function Symbol Characteristic Analog input 2 AIN2 19 - 30 VDC, 24mA Active/Passive (4 - 20 mA) Analog input 3 AIN3 19 - 30 VDC, 24mA Active/Passive (4 - 20 mA) Analog input 4 AIN4...
  • Page 55: Power Supply

    Connecting 6.2 Connecting SITRANS TDL ① Pressure sensor ② Electrical terminals ③ USB port ④ Ethernet port for web server connection ⑤ Add-on module interface (optional) ⑥ Front panel display connection 6.2.5 Power supply Specification Device requirements: • voltage range: 19 to 30 VDC, typ. 24 VDC •...
  • Page 56: Analog Outputs And Inputs

    Connecting 6.2 Connecting SITRANS TDL Power supply connection diagram Protection The device is protected by eFuse with integrated reverse-input polarity protection. The device can withstand positive and negative supply voltages up to ±60 VDC. In case of overvoltage, undervoltage, overcurrent or reverse-input polarity, the device is switched off and the red diode on the HOST board (at the 24VDC section) is illuminated.
  • Page 57 Analog input/output dip switches Analog outputs can operate in two modes accordingly to NAMUR NE43 and EN 15267-3:2007. SITRANS TDL generates the failure signal at the level of 3.3 mA. NAMUR NE43 provides guidelines for signaling failure information to the safety-interlock systems over a 4 to 20 mA loop.
  • Page 58: Active Analog Outputs

    Active analog outputs need no external power supply of the output terminal. SITRANS TDL provides power supply for the active analog outputs and thus analog outputs MUST NOT BE powered externally. Please set the dip switch to Active (A) before powering up the system.
  • Page 59: Passive Analog Outputs

    Active analog inputs need no external powering of the input terminal. Set analog input dip switch to A (active) before powering up the system. SITRANS TDL provides power supply for the active analog inputs and thus analog inputs MUST NOT BE powered externally.
  • Page 60: Passive Analog Inputs

    6.2.11 Digital outputs SITRANS TDL system provides 4 high side power switch (MOSFET) based digital outputs. Digital outputs are galvanically isolated from HOST board thus the digital outputs block must be powered externally (pin 21 and 22, 19-30 VDC). Each output can handle current up to 0.5 A. The status of each output is displayed on the HOST board with the blue LED in DOUT section.
  • Page 61: Resistance Temperature Detectors

      6.2.12 Resistance temperature detectors The SITRANS TDL system includes resistance temperature detectors (RTDs) input for measuring the process temperature. The system supports RTDs of PT100 and PT1000 type in 2-, 3-, and 4-wire connection scheme. Two dip switches are used to configure the system according to the actual RTD configuration.
  • Page 62 Connecting 6.2 Connecting SITRANS TDL The wiring configuration is set by using the dip switch next to terminals 11 and 12. Configuration Dip switch setting Connection diagram (next to terminals 11 and 12) 2-wire RTD 2-wire, 4-wire 3-wire RTD 3-wire 4-wire RTD...
  • Page 63: Loop Cable Connection

    (2 x 0.5 mm (20 AWG)) for power supply, synchronization and a grounding wire (green yellow). For proper operation, use only the loop cable supplied by Siemens. Please follow the step-by-step instructions for correct installation. ① Power supply and sychronization (twisted pair) ②...
  • Page 64 Connecting 6.2 Connecting SITRANS TDL   Grounding point for loop cable ① Grounding point After opening the receiv‐ er unit, you will see the following board. ① Fiber cable receptacle ② Receiver power supply Slide the loop cable through the gland placed on the bottom side of the receiver.
  • Page 65: Emitter Unit - Loop Cable Installation Instruction

    Connecting 6.2 Connecting SITRANS TDL 11. When the installation is complete, the assembly should look like on the following figure. Close the receiver unit lid.     Completed loop cable installation ① Grounding point 6.2.15 Emitter unit - loop cable installation instruction Place the other side of the loop cable in the gland and using two wrenches (size 24), tighten the connection.
  • Page 66: Customer Cable Connection

    Please use the cable with 0.5 mm (10 AWG) wires and external diameter between 7 to 12 mm (0.3 to 0.5 in) to fit into the gland. Cable supplied by Siemens is preferred. Please follow the step-by-step instruction for correct installation. SITRANS TDL...
  • Page 67: Customer Cable Installation Instruction

    Connecting 6.2 Connecting SITRANS TDL 6.2.17 Customer cable installation instruction 1. Cautiously dismount emitter unit housing lid by unscrewing each of the four bolts. 2. Place the sealing nut and the claw on the customer cable. Sealing nut and claw on the customer cable ①...
  • Page 68: Ethernet Connection

    Ethernet connection A separate ethernet cable is connected via special gland for communication with the analyzer via web server. To ensure IP65, use only the ethernet cable supplied by Siemens. Please follow the step-by-step instruction for correct ethernet cable installation (Page 69).
  • Page 69: Ethernet Cable Installation Instruction

    6.2.20 Industrial protocols connection The SITRANS TDL offers the optional to use MODBUS RTU, MODBUS TCP and PROFIBUS industrial communication protocols. NOTICE Do not connect to two different networks simultaneously The second MODBUS TCP port is only for redundancy or for a ring network configuration.
  • Page 70 NOTICE Network Security Solution It is recommended to provide a secure network environment for SITRANS TDL via special routers or gateways. Additionally, security can be provided by segmenting individual subnets, for example, by using the cell security concept or a demilitarized zone (DMZ).
  • Page 71: Commissioning

    The same risk continues to apply when error messages are switched off or disabled. WARNING Hot surfaces Risk of burns resulting from hot surfaces. • Take corresponding protective measures, for example by wearing protective gloves. SITRANS TDL Operating Instructions, 05/2023, A5E51429088_AA...
  • Page 72: General Requirements

    The device is built for easy operation, making it possible to be commissioned quickly. Its parameters are menu-driven and can be modified through local operation, using the display and buttons, also known as the Human Machine Interface (HMI). SITRANS TDL Operating Instructions, 05/2023, A5E51429088_AA...
  • Page 73: Device Startup

    During the start-up procedure, the LED will flash red until complete. Initialization takes approximately six minutes to complete. Device startup Note Output remains active While the device is being configured, the output remains active and continues to respond to changes in the process. SITRANS TDL Operating Instructions, 05/2023, A5E51429088_AA...
  • Page 74 Device parameters check and set Lasers temperature stabilization (laser off) Lasers temperature stabilization (laser on) Lasers transmission check Calibration check procedure Concentration measurement check Normal operation of the device (startup procedure is fin‐ ished) SITRANS TDL Operating Instructions, 05/2023, A5E51429088_AA...
  • Page 75: Access Control

    • If User PIN is lost, it can be reset with expert access level. • If Expert PIN is lost, see Recovery ID (5.3) (Page 132) under menu "Security" to reset the PIN with the help of Siemens customer support. SITRANS TDL...
  • Page 76: Local Commissioning

    Perform customization for your application only after "Quick start" wizard has been completed. Note Parameter settings properly stored In order to ensure parameter settings are properly stored in the device, wait 30 seconds after any configuration change before removing power from the device. SITRANS TDL Operating Instructions, 05/2023, A5E51429088_AA...
  • Page 77 Some of the parameters and settings shown in the operation instructions may not appear on the actual device, visibility is based on the application and configuration selected. Button Function Leave menu without saving changes Scroll up in list of options SITRANS TDL Operating Instructions, 05/2023, A5E51429088_AA...
  • Page 78: Quick Commissioning

    • Settings for quick commissioning wizard are inter-related and changes apply only after steps are finished. • Do not use quick commissioning wizard to modify individual parameters. (See instead chapter Parameter assignment (Page 113).) Perform customization for your application only after "Quick commissioning wizard" has been completed. SITRANS TDL Operating Instructions, 05/2023, A5E51429088_AA...
  • Page 79 While the device is being configured, the output remains active and continues to respond to changes in the process. The Quick commissioning wizard will guide you through configuration of parameters essential for your application. SITRANS TDL Operating Instructions, 05/2023, A5E51429088_AA...
  • Page 80 Commissioning 7.5 Local commissioning SITRANS TDL Operating Instructions, 05/2023, A5E51429088_AA...
  • Page 81: Operating

    Visibility of some menu/parameter items is based on configuration and access level. Although all visible items and their settings can be viewed in the HMI menu, parameters can be protected against changes through access level control. SITRANS TDL Operating Instructions, 05/2023, A5E51429088_AA...
  • Page 82: Display Views

    • If User PIN is lost, it can be reset with expert access level. • If Expert PIN is lost, see Recovery ID (5.3) (Page 132) under menu "Security" to reset the PIN with the help of Siemens customer support. Disable access level control If logged in as Expert, the user PIN can be disabled (parameter "Disable user PIN").
  • Page 83 Up to six user-configured views are available in operation view: Default views at initial startup are set as follows (exemplary for oxygen measuring variant of SITRANS TDL): • View 1: Concentration O (1 value) •...
  • Page 84 Parameter view can only be accessed by pressing the right button in a measurement view type. For more information, see menu Local display (2.8) (Page 122). View type: Measurement The operation view type "Measurement" shows measurement values. The following options can be selected: 1 value SITRANS TDL Operating Instructions, 05/2023, A5E51429088_AA...
  • Page 85 For more information, see menu Local display (2.8) (Page 122). Button functions: Measurement Button Function No functionality Go to the previous view in the operation view Go to the next view in the operation view Enter the parameter (navigation) view SITRANS TDL Operating Instructions, 05/2023, A5E51429088_AA...
  • Page 86 The time of the acknowledgement is shown in the history log. • Auto: The alarm/diagnostic is removed from the list when the cause is removed. Choose setting in parameter Acknowledge mode (2.4.1.5). SITRANS TDL Operating Instructions, 05/2023, A5E51429088_AA...
  • Page 87 Select item below in list; press and hold button to accelerate scrolling down the selection list Enter alarm view level 3 Alarm view - level 3 Button Function Enter alarm view level 2 No functionality No functionality No functionality SITRANS TDL Operating Instructions, 05/2023, A5E51429088_AA...
  • Page 88 2.5.4 View 1 ... In parameter view, the first level of the 5-level menu structure (entered by pressing right arrow from operation view), is standardized for all Siemens Measurement Intelligence devices and covers the following groups: Menu item Menu item name...
  • Page 89 Select the item above in the list; keep pressing the button to accelerate scrolling up the selection list. If the button is pressed when the top item is selected, the bottom item will be highlighted. SITRANS TDL Operating Instructions, 05/2023, A5E51429088_AA...
  • Page 90 Help text describing the parameter function. The help text appears if no button is pressed for three seconds. ⑤ Minimum value ⑥ Value to be edited ⑦ Escape without saving (frame around "ESC" is only shown when cursor is in left-most position) ⑧ Maximum value SITRANS TDL Operating Instructions, 05/2023, A5E51429088_AA...
  • Page 91 ASCII characters: select the next character in the alphabet. Select the next right position. If most right position is selected, confirm the change and exit the parameter edit view. Keep pressing the button to jump to the most right position. SITRANS TDL Operating Instructions, 05/2023, A5E51429088_AA...
  • Page 92: Process Values

    The display default value may vary depending on the process value selected. All process values are available on each current output channel: • Current output 1 • Current output 2 • Current output 3 • Current output 4 SITRANS TDL Operating Instructions, 05/2023, A5E51429088_AA...
  • Page 93 The alarm is triggered when the limit value rises above the Upper limit (1). The alarm is cleared when the limit value falls below the Upper limit (2). All limits can be used as a source for the digital output (Page 121). SITRANS TDL Operating Instructions, 05/2023, A5E51429088_AA...
  • Page 94: Inputs And Outputs

    • Current output 3 - signal, lower range value, upper range value, units/simulation value • Current output 4 - signal, lower range value, upper range value, units/simulation value The loop current scale for each current output is 4 ... 20 mA. SITRANS TDL Operating Instructions, 05/2023, A5E51429088_AA...
  • Page 95 * If the concentration reading drops below 2.5%, the mA output drops below 4 mA. ** If the concentration reading rises above 22.5%, the mA output rises above 20 mA. *** If fail-safe state occurs, the mA output will read 3.5 mA (NAMUR value for parameter "Lower fault current"). SITRANS TDL Operating Instructions, 05/2023, A5E51429088_AA...
  • Page 96: Digital Output

    Device status (3.2.1) (Page 128) 8.1.4 Diagnostics SITRANS TDL allows the user to enable/disable alarms and diagnostics, such as those related to process values, current inputs and outputs. It is important to note that a specific alarm/diagnostic must first be enabled before a fault notification of the same type will display, when a true fault actually occurs.
  • Page 97: Diagnostic Log

    PINs are used for these access levels. To enable PINs, change a PIN, or recover a lost PIN, see Security (5) (Page 131). Remote operation SITRANS TDL supports the following ways for operation via remote communications: • via ethernet and standard web browser • Modbus RTU (RS485) •...
  • Page 98: Modbus Rtu

    Note Permanent connection to the internet SITRANS TDL is not intended for permanent connection to the internet. For all types of remote operation, the security information (Page 13) must be considered. 8.2.2.1 Establishing communication with the device...
  • Page 99 Operating 8.2 Remote operation 3. Click Ethernet connection to access adapter settings. 4. Click “Change adapter options”. SITRANS TDL Operating Instructions, 05/2023, A5E51429088_AA...
  • Page 100 Operating 8.2 Remote operation 5. Open Ethernet status window. 6. Right-click to open Properties. 7. Select IPv4 Internet Protocol and click Properties. SITRANS TDL Operating Instructions, 05/2023, A5E51429088_AA...
  • Page 101: User Access Level

    10.Open an internet browser (preferably Mozilla Firefox). 11.Enter in 192.168.16.xx (IP of device specific model found in HMI, Parameter 4.1.1). 12.If the Ethernet connection is established, the SITRANS TDL web page application opens. If the SITRANS TDL web page does not launch, try: •...
  • Page 102: Self-Monitoring And Diagnosis Tab

    Shows status of subsystem components Out of range Process temperature and pressure, gas concen‐ tration, path length and ambient temperature range Maintenance Laser transmission status and reference gas level Startup Procedure Startup parameter evaluation SITRANS TDL Operating Instructions, 05/2023, A5E51429088_AA...
  • Page 103: Measurements Tab

    (log points). It is not recommended to exceed the limit of 100,000 points. Download CSV Downloads measurements chosen in the table and save it in a comma-separated (*.csv) for‐ mat. Show plot Functionality of plot window SITRANS TDL Operating Instructions, 05/2023, A5E51429088_AA...
  • Page 104 – Concentration (display gas concentration) – Laser optical transmission between emitter unit and receiver) • TEC 0 (Laser temperature) – Tec 0 (display measurement from first laser temperature control module) • SYSTEM – Display system health SITRANS TDL Operating Instructions, 05/2023, A5E51429088_AA...
  • Page 105: Parameters Tab

    – Ain (Analog input measurement) • DigitalInOut The following sections describe how to set the measurement groups above using the SITRANS TDL web server. 8.2.2.5 Parameters tab The parameter tab allows configuring of external temperature or pressure signals fed in through the analog ports, as well as configuring analog output signals from the device.
  • Page 106 – Damping value T90 [s] • IO – Dout (Digital output 1 to 4) – Aout (Analog output configuration - Signal, Simulation value, Lower range value, Upper range value) – Ain (Analog input configuration) SITRANS TDL Operating Instructions, 05/2023, A5E51429088_AA...
  • Page 107 • Medium pressure / Source • Medium pressure / Fixed value [mbar] • Path length [m] • Damping value T90 [s] Display (Measurements/Show grid) Digital output window Options • Digital outputs 1 to 4 SITRANS TDL Operating Instructions, 05/2023, A5E51429088_AA...
  • Page 108 • Current outputs 1 to 4 – Signal – Simulation value – Lower range value – Upper range value Analog value is scaled to a defined range and set to the specific input. Display (Measurements/Show grid) SITRANS TDL Operating Instructions, 05/2023, A5E51429088_AA...
  • Page 109 • Current date and time / Month • Current date and time / Day • Current date and time / Hour • Current date and time / Minute • Current date and time / Second SITRANS TDL Operating Instructions, 05/2023, A5E51429088_AA...
  • Page 110: Settings Tab

    Attention: Use LOAD button only if you are sure with you want to do. This option return device to factory settings. All user settings will be lost. Note Save user settings There is the option to save user settings before loading factory configuration. SITRANS TDL Operating Instructions, 05/2023, A5E51429088_AA...
  • Page 111: About Tab

    Operating 8.2 Remote operation 8.2.2.8 About tab This page displays device type, firmware version and serial number. SITRANS TDL Operating Instructions, 05/2023, A5E51429088_AA...
  • Page 112 Operating 8.2 Remote operation SITRANS TDL Operating Instructions, 05/2023, A5E51429088_AA...
  • Page 113: Parameter Assignment

    In the device menu that follows, alarms and diagnostics that are enabled by default are denoted by an "X" beside the specific alarm/diagnostic name. Note Output remains active While the device is being configured, the output remains active and continues to respond to changes in the process. SITRANS TDL Operating Instructions, 05/2023, A5E51429088_AA...
  • Page 114: Quick Start (1)

    Lower range value available only if Current input 1 to 4 is chosen as source. Upper range value Sets process value that corresponds to 20 mA loop current. Note Upper range availability Upper range value available only if Current input 1 to 4 is chosen as source. SITRANS TDL Operating Instructions, 05/2023, A5E51429088_AA...
  • Page 115: Medium Temperature

    Sets process value that corresponds to 20 mA loop current. Note Upper range availability Upper range value available only if Current input 1 to 4 is chosen as source. Current value Displays the current temperature. SITRANS TDL Operating Instructions, 05/2023, A5E51429088_AA...
  • Page 116: Path Length

    Sets the configurable manual value if the process conditions are static. 9.2.1.3 Current value (2.1.3) Displays the current pressure. 9.2.2 Medium temperature (2.2) 9.2.2.1 Source (2.2.1) Sets the signal source the analyzer is using to collect the temperature of gas to be analyzed. SITRANS TDL Operating Instructions, 05/2023, A5E51429088_AA...
  • Page 117: Fixed Value (2.2.2)

    If the limit is not monitored by the HMI, web server or fieldbus, it is a value stored in the analyzer, but not assigned to a digital output. 9.2.4.1 Limit 1 (2.4.1) Enable (2.4.1.1) Enables/disables monitor values from the device. Setting • Disabled • Enabled SITRANS TDL Operating Instructions, 05/2023, A5E51429088_AA...
  • Page 118 Defines the level which the output will have if the limit condition is not fulfilled. Setting • Active low level • Active high level Hysteresis (2.4.1.7) Sets hysteresis limits. Hysteresis is distance between limits for activation and deactivation of an alarm/warning. SITRANS TDL Operating Instructions, 05/2023, A5E51429088_AA...
  • Page 119: Limit 2 (2.4.2) To Limit 10 (2.4.10)

    Current input linked to process parameter If a current input is linked to a process parameter, the corresponding unit is shown. Current input 1 (2.6.1.1) Lower range value (2.6.1.1.1) Sets process value that corresponds to 4 mA loop current. SITRANS TDL Operating Instructions, 05/2023, A5E51429088_AA...
  • Page 120: Current Output (2.6.2)

    Current output 2 (2.6.2.2) to Current output 4 (2.6.2.4) Note Parameters for "Current output 2" to "Current output 4" Parameters for "Current output 2" to "Current output 4" are the same as those listed previously (for "Current output 1": 2.6.2.1). SITRANS TDL Operating Instructions, 05/2023, A5E51429088_AA...
  • Page 121: Digital Output (2.6.3)

    Set date and time (2.7.2) Sets the current date in the format YYYY-MM-DD and the time in 24-hour format. For information on how to use the string editor, see Edit view/Read only view (Page 90). SITRANS TDL Operating Instructions, 05/2023, A5E51429088_AA...
  • Page 122: Local Display (2.8)

    Always on The display backlight will turn on with first button press, and turn off 30 seconds after last button press when set to "Auto". 9.2.8.3 Contrast (2.8.3) Sets contrast level on local display. SITRANS TDL Operating Instructions, 05/2023, A5E51429088_AA...
  • Page 123: View 1 (2.8.4)

    • Concentration O • Laser transmission • Fiber transmission • Reference amplitude 2nd value (2.8.4.3) Sets the process value to show on local display based on number of process values selected in view type. SITRANS TDL Operating Instructions, 05/2023, A5E51429088_AA...
  • Page 124 Sets the process value to show on local display based on number of process values selected in view type. Setting • Medium temperature • Medium pressure • Concentration O • Fiber transmission • Laser transmission • Reference amplitude SITRANS TDL Operating Instructions, 05/2023, A5E51429088_AA...
  • Page 125: View 2 (2.8.5)

    Sets upper limit of axis for graph on local display view. Graph scale lower limit (2.8.4.11) Sets lower limit of axis for graph on local display view. 9.2.8.5 View 2 (2.8.5) Setting • Disabled • Enabled SITRANS TDL Operating Instructions, 05/2023, A5E51429088_AA...
  • Page 126 • • Graph scale upper limit (2.8.5.11) Sets upper limit of axis for graph on local display view. Graph scale lower limit (2.8.5.12) Sets lower limit of axis for graph on local display view. SITRANS TDL Operating Instructions, 05/2023, A5E51429088_AA...
  • Page 127: View 3 (2.8.5) To View 6 (2.8.9)

    Displays version number corresponding to local display of device. FW version (3.1.7.2) Displays version number corresponding to software or firmware embedded in local display. Menu structure version (3.1.7.3) Displays the current version number of the HMI firmware. SITRANS TDL Operating Instructions, 05/2023, A5E51429088_AA...
  • Page 128: Diagnostics (3.2)

    Displays amplification of the signal received in the receiver. Reference amplitude (3.3.1.4) Displays amplitude of the measurement line of the reference gas. Laser temperature (3.3.1.5) Displays current temperature of the laser (this changes over time and is an indicator of laser aging). SITRANS TDL Operating Instructions, 05/2023, A5E51429088_AA...
  • Page 129: Process Values (3.3.2)

    Parameters for "Current input 1" to "Current input 4" Parameters for "Current input 1" to "Current input 4" are the same (for "Current input": 3.3.3.1 to 3.3.3.4). Specific diagnostic codes are available here: Diagnostics and troubleshooting (Page 139). SITRANS TDL Operating Instructions, 05/2023, A5E51429088_AA...
  • Page 130: Outputs (3.3.4)

    Used to confirm if memory card installed in device, and if it is being used as a mass storage device (MSD). Memory card includes backup data for device. Setting • • • Used as mass storage device • Checking memory card... 9.3.4.2 Capacity (3.7.3) Displays total capacity of installed memory card. SITRANS TDL Operating Instructions, 05/2023, A5E51429088_AA...
  • Page 131: Free Space (3.7.4)

    Used to change PIN code that enables user access level. 9.5.2 Change expert PIN (5.2) Used to change PIN code that enables expert access level. This parameter is only visible with "Expert" access level. SITRANS TDL Operating Instructions, 05/2023, A5E51429088_AA...
  • Page 132: Recovery Id (5.3)

    (Serial number can be found on device nameplate.) When you provide this information to Siemens Technical Support, a PUK (PIN Unlock Key) will be supplied to you. Enter this PUK in parameter "PIN recovery" to reset the user PIN to factory default value.
  • Page 133: Service And Maintenance

    • Remove dust layers in excess of 5 mm. CAUTION Releasing button lock Improper modification of parameters could influence process safety. • Make sure that only authorized personnel may cancel the button locking of devices for safety-related applications. SITRANS TDL Operating Instructions, 05/2023, A5E51429088_AA...
  • Page 134: Cleaning

    • Repair must be carried out by Siemens authorized personnel only. WARNING Impermissible repair and maintenance of the device • Repair and maintenance must be carried out by Siemens authorized personnel only. WARNING Impermissible accessories and spare parts • Only use original accessories or original spare parts.
  • Page 135 • As long as the device is under pressure, do not loosen process connections and do not remove any parts that are pressurized. • Before opening or removing the device ensure that process media cannot be released. SITRANS TDL Operating Instructions, 05/2023, A5E51429088_AA...
  • Page 136: Replacing The Device

    • Make sure that the sensor enclosure is not damaged when replacing the sensor. • If contact with the corrosive substances occurs, rinse the affected skin immediately with large amount of water to dilute substance. SITRANS TDL Operating Instructions, 05/2023, A5E51429088_AA...
  • Page 137: Transport

    To return a product to Siemens, see Returns to Siemens (www.siemens.com/returns-to- siemens). Contact your Siemens representative to clarify if a product is repairable, and how to return it. They can also help with quick repair processing, a repair cost estimate, or a repair report/ cause of failure report.
  • Page 138 Service and maintenance 10.7 Disposal Note Special disposal required The device includes components that require special disposal. • Dispose of the device properly and environmentally through a local waste disposal contractor. SITRANS TDL Operating Instructions, 05/2023, A5E51429088_AA...
  • Page 139: Diagnostics And Troubleshooting

    11.1 Calibration information The SITRANS TDL system is factory calibrated using certified gas mixtures that can be traced to national or international standards . The device uses an internal in-line reference gas cell for real time verification of the calibration status or a process gas normally present in the process gas stream that is used to perform permanent self supervision of the analyzer.
  • Page 140 Diagnostics and troubleshooting 11.2 Device status symbols SITRANS TDL Operating Instructions, 05/2023, A5E51429088_AA...
  • Page 141: Technical Specifications

    Technical specifications Note Device specifications Siemens makes every attempt to ensure the accuracy of these specifications but reserves the right to change them at any time. Note Device-specific approvals Always refer to nameplates on the device for device-specific approvals. CAUTION...
  • Page 142: Power

    < 15VA EMC immunity In accordance with EN 61326-1 Electric safety In accordance with 61010-1 12.3 Performance Warm-up time: approx. 6 minutes. It takes approximately 6 minutes for the system to be fully operational. SITRANS TDL Operating Instructions, 05/2023, A5E51429088_AA...
  • Page 143: Inputs/Outputs

    If these conditiions cannot be fulfilled, a heat protection must be installed. Depending on local conditions, solutions may range from installation of a sun shield to operating SITRANS TDL in a cabinet with controlled environment.
  • Page 144: Packaging

    Documents 1. Packing list 2. Calibration certificate 3. Analyzer test report Tools 1. Wrench size 24 2. Hex 6 mm (0.2 in) 3. Hex 14 mm (0.6 in) 4. Flat-head screwdriver (for electrical connections) SITRANS TDL Operating Instructions, 05/2023, A5E51429088_AA...
  • Page 145: Approvals Non-Hazardous

    Technical specifications 12.9 Approvals non-hazardous 12.9 Approvals non-hazardous Note The device nameplate lists the approvals that apply to your device. General purpose SITRANS TDL Operating Instructions, 05/2023, A5E51429088_AA...
  • Page 146 Technical specifications 12.9 Approvals non-hazardous SITRANS TDL Operating Instructions, 05/2023, A5E51429088_AA...
  • Page 147: Dimension Drawings

    Dimension drawings 13.1 Emitter and receiver dimensions Note Sufficient space necessary to operate the display To view and operate the display, allow sufficient space around the emitter. SITRANS TDL Operating Instructions, 05/2023, A5E51429088_AA...
  • Page 148: Dimension Drawings

    Dimension drawings 13.2 Dimension drawings 13.2 Dimension drawings SITRANS TDL (emitter and receiver) ① Purging tube ② Process flange ③ Process interface ④ 6° Maximum Dimensions in mm (inch) Emitter unit Dimensions in mm (inch)   Receiver unit Dimension in mm (inch)  ...
  • Page 149 Dimension drawings 13.2 Dimension drawings Customer flange DN50/PN10-40, all measures in mm (inches) Dimension in mm (inch)   SITRANS TDL Operating Instructions, 05/2023, A5E51429088_AA...
  • Page 150 Dimension drawings 13.2 Dimension drawings Customer flange ANSI 4"/150 lb, all measures in mm (inches) Dimension in mm (inch)   SITRANS TDL Operating Instructions, 05/2023, A5E51429088_AA...
  • Page 151: Product Documentation And Support

    – "Download as html5, only PC": Open or save the manual in the HTML5 view on your PC You can also find manuals with the Mobile app at Industry Online Support (https:// support.industry.siemens.com/cs/ww/en/sc/2067). Download the app to your mobile device and scan the device QR code. Product documentation by serial number Using the PIA Life Cycle Portal, you can access the serial number-specific product information including technical specifications, spare parts, calibration data, or factory certificates.
  • Page 152: Technical Support

    (http:// www.siemens.com/automation/csi/service).   Service & support on the Internet In addition to our technical support, Siemens offers comprehensive online services at service & support (http://www.siemens.com/automation/serviceandsupport). Contact If you have further questions about the device, contact your local Siemens representative at Personal Contact (http://www.automation.siemens.com/partner).
  • Page 153: Technical Reference

    This quantity is estimated for each application and is determined for each device before shipping to our customers. For every SITRANS TDL analyzer, we determine the limit of detection during a 12-hour cycle in a climate chamber. The analyzer is placed on a test gas cell and a zero sample is introduced.
  • Page 154: Diagnostic Leds

    – 0.5 Hz flashing LED when product has an active alarm for process alarm/warning or maintenance demanded/required • Yellow LED – When device is responding to a field bus request that was addressed to it  SITRANS TDL Operating Instructions, 05/2023, A5E51429088_AA...
  • Page 155: Communication

    C.2.2 PROFIBUS communication SITRANS TDL with the addition of a PROFIBUS DP communication card allows communication with a PROFIBUS DP leader, for cyclic communication. SITRANS TDL Operating Instructions, 05/2023, A5E51429088_AA...
  • Page 156 Communication C.2 PROFIBUS SITRANS TDL Operating Instructions, 05/2023, A5E51429088_AA...
  • Page 157: D Remote Operation

    Remote operation Modbus RTU (RS485) Modbus RTU is fully implemented by Siemens and does not need additional communication module. Device can work in follower mode. Communication parameters are configured by SITRANS TDL Web server. ① Location RS-485 socket on host PCB  ...
  • Page 158: Simatic Pdm

    Operating Manual 'Help for SIMATIC PDM'. The manual is delivered with SIMATIC PDM software. Once the SIMATIC PDM is installed on your computer you find the manual under: Start > All programs > Siemens Automation > SIMATIC > Documentation. Link at our website: SIMATIC PDM instructions and manuals (https:// support.industry.siemens.com/cs/ww/en/ps/16983/man).
  • Page 159: Updating The Electronic Device Description (Edd) Or Field Device Integration (Fdi)

    – The most recent Service Pack (SP) – The most recent hot fix (HF) Note Modbus RTU for SIMATIC PDM connection It is recommended to use Modbus RTU for connection of SITRANS TDL to SIMATIC PDM. SITRANS TDL Operating Instructions, 05/2023, A5E51429088_AA...
  • Page 160 Remote operation D.3 SIMATIC PDM SITRANS TDL Operating Instructions, 05/2023, A5E51429088_AA...
  • Page 161: Index

    Laws and directives for Korea, 16 Scope of delivery, 12 HMI (Human Machine Interface), 72, 81 Service, 152 Hotline, (Refer to Support request) Service and support Internet, 152 Support, 152 Support request, 152 Symbols, (Refer to warning symbols) Laws and directives, 16 Lithium batteries, 137 Technical data, 141 SITRANS TDL Operating Instructions, 05/2023, A5E51429088_AA...
  • Page 162 Index Technical support, 152 partner, 152 personal contact, 152 Test certificates, 15 Views access level, 90 alarm, 86 diagnostics, 86 edit, 90 operation, 83 parameter, 88 read only, 90 Warning symbols, 15 Warranty, 14 SITRANS TDL Operating Instructions, 05/2023, A5E51429088_AA...