Safety notes

3.2 Requirements for special applications

3.2
Requirements for special applications
Due to the large number of possible applications, each detail of the described device versions
for each possible scenario during commissioning, operation, maintenance or operation in
systems cannot be considered in the instructions. If you need additional information not
covered by these instructions, contact your local Siemens office or company representative.
Note
Operation under special ambient conditions
We highly recommend that you contact your Siemens representative or our application
department before you operate the device under special ambient conditions as can be
encountered in nuclear power plants or when the device is used for research and
development purposes.
Pressure safety
For pressure specifications, refer to the Technical specifications (Page 52).
Damage to operating staff
Whenever the optics housings heads are to be opened or disconnected from the process
interface or the gas tubing, make sure that any overpressure in the optics housings heads is
removed prior to any work. Opening optics housings under overpressure may result in
dangerous injuries to the operator.
Damaged optics
The high pressure-proof glass in the optics housings must not be subject to any mechanical
impact which might cause damage to the glass (scratch, gash etc.). Only use a soft cloth for
cleaning of the glass. Make sure that it is safe to dismount the optics housing before
cleaning.
Malfunction of sealing
Make sure all plugs and tube fittings are sealed with the enclosed pipe thread sealant.
Instructions given on the sealant tube must be carefully followed.
10
WARNING
WARNING
WARNING
Compact Operating Instructions, 05\2023, A5E51659183-AA – revised 03/2024
SITRANS TDL