Commissioning

7.1

Basic safety notes

Toxic gases and liquids
Danger of poisoning when venting the device: if toxic process media are measured, toxic
gases and liquids can be released.
• Before venting ensure that there are no toxic gases or liquids in the device, or take the
appropriate safety measures.
Improper commissioning
Device failure
• Do not commission the device until it has been mounted completely and connected in
accordance with the information in Technical specifications (Page 52).
• Before commissioning, take the effect on other devices in the system into account.
Commissioning and operation with error message
If an error message displays, correct operation is no longer guaranteed.
• Check the severity of the error.
• Correct the error.
• If the error still exists:
– Take the device out of operation.
– Do not restart the device.
The same risk continues to apply when error messages are switched off or disabled.
Hot surfaces
Risk of burns resulting from hot surfaces.
• Take corresponding protective measures, for example by wearing protective gloves.
Compact Operating Instructions, 05\2023, A5E51659183-AA – revised 03/2024
DANGER
WARNING
WARNING
WARNING
SITRANS TDL
7
39