Emitter Unit - Loop Cable Installation Instruction - Siemens SITRANS TDL Compact Operating Instructions

In situ laser gas analyzers
Hide thumbs Also See for SITRANS TDL:
Table of Contents

Advertisement

6. Place two black wires in sockets
of number 25 and 26 for power-
ing the analyzer. The same wires
should be then connected to the
24VDC power supply on the cus-
tomer side.
7. Place other wires in proper sock-
ets for connection with analog
and digital inputs/outputs as
desired for specific application of
the analyzer. Connect with ana-
log and digital inputs/outputs as
desired for specific application of
the analyzer.
6.2.7

Emitter unit - loop cable installation instruction

1.
Place the other side of the loop cable in the gland and using two wrenches (size 24), tighten the connec-
tion. The loop cable should now be blocked in the gland. This can be checked by gently pulling the loop
cable.
2.
Place the fiber optic cable in the
connector. Lock the cable by twist-
ing the nut on the thread.
SITRANS TDL
Compact Operating Instructions, 05\2023, A5E51659183-AA – revised 03/2024
Powering the analyzer
Connection with analog and digital inputs/outputs
① Fiber cable connector (emitter unit)
Connecting
6.2 Connecting SITRANS TDL
37