Installation guidelines

4.3 Alignment

4.2

Electrical connections

Installation of transmitter optics housing
Install the cables, hybrid and sensor connection cable, at the transmitter optics housing
according to Figure 3.2. Cables are fixed in the optics housing with gaskets. Always use lubricant
on the gaskets before installing, otherwise these will be very difficult to remove later. Multi-
mode fiber (SMA connector) and 24 V are just connected straight through to the receiver.
Note
Always protect the single-mode fiber (E2000 connector) from dust. Do not remove the plastic
vial until the cable is firmly fastened in the optics housing.
Figure 4-2
Installation of receiver optics housing
Install the sensor connection cable in the receiver optics housing. Cables are fixed with gaskets.
Always use lubricant on the gaskets before installing, otherwise they will be very difficult to
remove later.
Connect the multi-mode fiber to the sensor-board as shown below.
Figure 4-3
4.3
Alignment
Mount the optics housings, receiver and transmitter, with gaskets on the flanges and cross
tighten the bolts.
When mounting the optics housings the large springs of the alignment flanges must be located
in the lower section of the flange. All purging outlets must face downwards.
20
Connections in the transmitter
Connections in the receiver
LDS 6, Optics Housing CD 6
Operating Instructions, 01/2022, A5E01134377-05