Fluke 5730A Operator's Manual page 72

Multifunction calibrator
Hide thumbs Also See for 5730A:
Table of Contents

Advertisement

5730A
Operators Manual
The 10.000208 V output setting is calculated by the Calibrator with the subsequent
equation:
This is applied to 10 V to yields:
The 1.3 mV zero offset is then subtracted:
8. Use the output adjustment controls to adjust the Calibrator output for a reading of
The display now shows that the DMM scale error at 19.9 V is -0.0151 %, and the DMM
linearity error at 10 V is +0.0352 %.
4-44
19.9 V – 19.903 V
1 -
10 V x 1.0001508 = 10.001508 V
10.001508 V − 0.0013 V = 10.000208 V
10.0 V (the reference value) on the DMM. The displays change to:
= 1.0001508
19.9 V
hhp141.tif