Advertisement

5730A
Operators Manual
*WAI
Description
Parameter
Example
WBAND
Description
Parameter
XFER
Description
Parameter
Example
XFER?
Description
Parameter
Example
XFERCHOICE
Description
Parameter
Example:
XFERCHOICE?
Description
Parameter
Response
6-46
Table 6-5. Commands (cont.)
Sequential command.
(Wait-to-Continue.) This command prevents further remote commands from being
executed until all previous remote commands have been executed. (See also *OPC.)
None
If an OUT command had been sent, it can cause the Calibrator to wait until the output
has settled before continuing on to the next command by following OUT with a *WAI
command. This is useful because OUT is an overlapped command, which means the
Calibrator would normally go on to process other commands before completing the OUT
command.
Overlapped command, ignored if not in remote.
Activates or deactivates output from the Wideband AC Module (Option 5730A/03). This
has the same action as touching the wideband selection from the front panel during local
operation.
ON or OFF
Turns ac/dc transfers off (or on) after output has settled.
OFF or ON
XFER OFF turns transfers off (ON is the default)
Returns whether ac/dc transfers after settling are turned on.
None
ON or OFF
Sets whether the choice to turn ac/dc transfers off is given on the front panel GUI for
ac V outputs.
ON or OFF
XFERCHOICE OFF omits the choice from the front panel (this is the default setting).
Returns whether the choice to turn ac/dc transfers off is given on the front panel GUI for
ac V outputs.
None
ON or OFF