Fluke 5730A Operator's Manual page 140

Multifunction calibrator
Hide thumbs Also See for 5730A:
Table of Contents

Advertisement

5730A
Operators Manual
SP_EOF
Description
Parameter
SP_EOF?
Description
Parameter
Response
SP_SET
Description
Parameter
Default
Example
6-42
Table 6-5. Commands (cont.)
Sequential command, ignored if not in remote.
Sets the End-Of-File character string for the RS-232-C serial interface. The EOF setting
is saved when the power is turned off.
1.
ASCII code (decimal) for the first character
2.
ASCII code (decimal) for the second character
(A "0" for parameter 1 or 2 means that there is no character in that position.)
Sequential command.
Returns the End-Of-File character string.
None
1.
(Integer) ASCII code in decimal for the 1st character
2.
(Integer) ASCII code in decimal for the 2nd character
(A "0" for response 1 or 2 means there is no character in that position.)
Sequential command, ignored if not in remote.
Sets the serial port settings and saves them in nonvolatile memory.
One or more of the following, in any order:
1.
Baud rate as 9600, 19200, 38400, 57600 or 115200
2.
TERM or COMP
3.
XON, RTS, or NOSTALL
4.
DBIT7 or DBIT8
5.
SBIT1 or SBIT2
6.
PNONE, PEVEN, PODD, or PIGNORE
7.
CR, LF, or CRLF
The TERM or COMP parameter sets the responses in serial
remote control to be appropriate for interactive terminal use
or operation under program control. Specifying TERM sets
the remote port to expect a human operator using a terminal.
Specifying COMP sets the remote port to expect a computer
program to be controlling the Calibrator. (This command
has the same effect as setting Remote I/F to TERMINAL or
COMPUTER in the RS-232C Port Setup Menu.)
9600,TERM,XON,DBIT8,SBIT1,PNONE,CRLF
19200,COMP,XON,DBIT8,SBIT1,PEVEN,CRLF
(See note below)
(Stall method)
(Data bits)
(Stop bits)
(Parity)
(End-Of-Line)
Note