Checking Pressure Accuracy

You can check the pressure accuracy of the device.
[Power ON/OFF]
button
[Auscultation]
button
1. Make sure the power is on.
"Zero setting" is performed.
When "zero setting" is finished, the pressure accuracy verification
screen appears.
the pressure accuracy check.
Apply the external pressure.
Compare with the displayed value and
make sure there is no problem.
Example:
1. Connect the blood pressure monitor,
the calibrated reference pressure
gauge, and the cuff and inflation bulb.
Note:
• Make sure that the blood pressure monitor reading is within ±3 mmHg
compared to the calibrated reference pressure gauge.
4. Turn off the power and exit.