Omron HEM-7280T-E Instruction Manual

Omron HEM-7280T-E Instruction Manual

Automatic upper arm blood pressure monitor

Advertisement

Automatic Upper Arm Blood Pressure Monitor
MIT5s Connect (HEM-7280T-E)
Instruction Manual
IM-HEM-7280T-E-03-02/2017
5344802-4C

Advertisement

Table of Contents
loading

Summary of Contents for Omron HEM-7280T-E

  • Page 1 Automatic Upper Arm Blood Pressure Monitor MIT5s Connect (HEM-7280T-E) Instruction Manual IM-HEM-7280T-E-03-02/2017 5344802-4C...
  • Page 3 H A A H AB...
  • Page 4: Table Of Contents

    Contents Thank you for purchasing the OMRON MIT5s Connect Automatic Upper Arm Blood Pressure Monitor. Your new blood pressure monitor uses the oscillometric method of blood pressure measurement. This means the monitor detects your blood movement through your brachial artery and converts the movements into a digital reading. An oscillometric monitor does not need a stethoscope so the monitor is simple to use.
  • Page 5: Important Safety Information

    Important Safety Information Warning: Indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury. (General Usage) DO NOT adjust medication based on measurement results from this blood pressure monitor. Take medication as prescribed by your physician. Only a physician is qualified to diagnose and treat High Blood Pressure. Consult your physician before using the device for any of the following conditions: common arrhythmias such as atrial or ventricular premature beats or atrial fibrillation, arterial sclerosis, poor perfusion, diabetes, pregnancy, pre-eclampsia, renal diseases.
  • Page 6 Important Safety Information Caution: Indicates a potentially hazardous situation which, if not avoided, may result in minor or moderate injury to the user or patient or damage to the equipment or other property. (General Usage) Always consult your physician. Self-diagnosis of measurement results and self-treatment are dangerous. Consult your physician before using the device for any of the following conditions: •...
  • Page 7 Wipe the dust off from the power plug. Unplug monitor when not in use. Disconnect the power plug before cleaning. Use only an OMRON AC adapter designed for this device. Use of unsupported adapters may damage and/or may be hazardous to the device. (Battery Usage) Do not insert the batteries with their polarities incorrectly aligned.
  • Page 8: Know Your Device

    1. Know Your Device Open the rear cover page to read the following: The letter identifiers on the rear cover page correspond to those in the body of this page. Monitor USER ID symbol Display Average value symbol Connection button Systolic blood pressure Up/Down buttons Diastolic blood pressure...
  • Page 9: Display Symbols

    1. Know Your Device 1.1 Display Symbols Irregular Heartbeat Symbol ( Normal Heartbeat When the monitor detects an irregular rhythm two or more times during the Pulse measurement, the irregular heartbeat symbol will appear on the display with Blood pressure the measurement values.
  • Page 10: Before Taking A Measurement

    1. Know Your Device Blood Pressure Level Indicator (Colour) If your Systolic Blood Pressure is 135 mmHg or above and/or the Diastolic Blood Pressure is 85 mmHg or above, the blood pressure level indicator (colour) will light in “orange” when the measurement result is displayed.
  • Page 11: Preparation

    2. Preparation 2.1 Battery Installation Remove the battery cover. Insert 4 “AA” batteries as indicated in the battery compartment. Replace the battery cover. Notes: • Please keep the batteries in your device all the time when you use an optional AC adapter. •...
  • Page 12: Using This Device With A Smartphone

    2. Preparation 2.2 Using this device with a Smartphone To begin using the OMRON connect app for the first time, please visit www.omronconnect.com/setup for the initial set-up instructions. List of Compatible Smartphones at: www.omronconnect.com/devices For details and more information, visit: www.omronconnect.com Download and install the free “OMRON connect”...
  • Page 13 2. Preparation Continue the pairing instructions shown in the OMRON connect app, while the connection symbol is flashing on the device’s display. Confirm if the device is connected successfully. When the device is connected successfully to the smartphone, the “OK”...
  • Page 14: Setting The Date And Time Manually

    2. Preparation 2.3 Setting the Date and Time Manually The date and time will be automatically set when paired with your smartphone. To set manually, follow the instructions below. Press the connection button. 1If the device has not been paired with a smartphone, the year flashes on the Date/Time display.
  • Page 15 2. Preparation Push or button to change the year. Push button to confirm the year, once confirmed the month starts flashing. Repeat the same steps to change the month, day, hour, and minute. Year Month Hour Change Confirm Minute Change Confirm : Forward : Back...
  • Page 16: Using The Device

    • The blood pressure can differ between the right arm and the left arm, and the measured blood pressure values can be different. OMRON recommends to always use the same arm for measurement. If the values between both arms differ substantially, please check with your physician which arm to use for your...
  • Page 17: How To Sit Correctly

    3. Using the Device 3.2 How to Sit Correctly To take a measurement, you need to be relaxed and comfortably seated, under comfortable room temperature. Avoid bathing, drinking alcohol or caffeine, smoking, exercising or eating 30 minutes before taking a measurement. •...
  • Page 18: Taking A Measurement

    3. Using the Device 3.3 Taking a Measurement Check if the date and time are correct on the display before taking the measurement. If the date and time settings are necessary, refer to “Setting the Date and Time Manually” (page 10). Notes: •...
  • Page 19: Using The Guest Mode

    3. Using the Device Remove the arm cuff. Press the START/STOP button to turn the monitor off. The monitor automatically stores the measurement result in its memory. It will automatically turn off after 2 minutes without any operations. Note: Wait 2-3 minutes before taking another measurement. Waiting between measurements allows the arteries to return to the condition prior to taking a measurement.
  • Page 20: Using The Memory Function

    3. Using the Device 3.4 Using the Memory Function The monitor automatically stores the results up to 100 sets for each user (1 and 2). It can also calculate an average value based on the last 3 measurement values taken within 10 minutes.
  • Page 21 3. Using the Device To View the Average Value Select your USER ID (1 or 2). Press and hold the button for more than 3 seconds. Notes: • If the previous measurement was taken without setting the date and time, the average value is not calculated.
  • Page 22 3. Using the Device Press the button. The morning average for “THIS WEEK” appears on the display. Note: The morning hypertension symbol ( ) appears if the morning weekly average is 135/85 mmHg or above. Press the button again, and the evening average for “THIS WEEK” appears on the display.
  • Page 23 3. Using the Device Morning Weekly Average Measurements of the morning This is the average for the Sunday Monday Tuesday Wednesday Thursday Friday Saturday 4:00 measurements taken during the morning First measurement of Measurement (4:00 - 11:59) between Sunday and the morning Measurement Up to 3 measurements...
  • Page 24 *Sync occurs if the time between smartphone and device differs more than 10 minutes. Open the OMRON connect app on your smartphone. Pull down to refresh the screen of the OMRON connect app and the data transfer starts. * If the data transfer is not starting, refer to “Connection failure. Data cannot be sent.”...
  • Page 25 3. Using the Device TO DELETE ALL RESULTS AND SETTINGS To delete all the information stored in your monitor, follow the instructions below. Make sure that the monitor’s power has been off. While holding the button down, press the START/STOP button for more than 5 seconds.
  • Page 26: Deleting All Values Stored In Memory

    3. Using the Device To Delete All the Values Stored in Memory The values stored in the memory are deleted by USER ID. Select your USER ID (1 or 2). Press the Memory button, while the memory symbol ( ) appears.
  • Page 27: Turning "Off" The Bluetooth

    3. Using the Device 3.5 Turning “OFF” the Bluetooth ® Turn off the Bluetooth ® of the monitor in the following areas where use of wireless equipment is prohibited. • On aircraft • In hospitals •While abroad Make sure that the power has been off before starting the instructions below. Press and hold the connection button for more than 10 seconds.
  • Page 28 3. Using the Device To turn “ON” the Bluetooth ® Make sure that the power has been off before starting the instructions below. Press and hold the connection button for more than 10 seconds. Release the button when “ON” appears on the display. Press the START/STOP button to turn the monitor off.
  • Page 29: Error Messages And Troubleshooting

    4. Error Messages and Troubleshooting 4.1 Error Messages Display Cause Solution Remove the arm cuff. Wait 2-3 minutes and then take another measurement. Irregular heartbeats are detected. Repeat the steps in section 3.3. If this error continues to appear, contact your physician. Apply the arm cuff tighter.
  • Page 30 Repeat measurement. Remain still and do not talk during measurement. Movement during measurement. Refer to section 3.3. Remove any clothing interfering with the arm cuff. Clothing is interfering with the arm cuff. Refer to section 3.1. Contact your OMRON retail outlet or Device error. distributor.
  • Page 31: Troubleshooting

    4. Error Messages and Troubleshooting 4.2 Troubleshooting Problem Cause Solution Apply the arm cuff tighter. Arm cuff is applied too loosely. Refer to section 3.1. Remain still and do not talk during Movement or talking during measurement. The measurement result is extremely measurement.
  • Page 32 If the Err symbol still lights after checking the application, contact customer service. • Press the START/STOP button and repeat measurement. • Replace the batteries with new ones. Other problems. If the problem continues, contact your OMRON retail outlet or distributor.
  • Page 33: Maintenance And Storage

    • Use a soft and dry cloth, or a soft and moistened cloth and neutral soap to clean on the monitor and the arm cuff. • Changes or modification not approved by the manufacturer will void the user warranty. Do not disassemble or attempt to repair the device or components. Consult your OMRON retail outlet or distributor. Calibration and Service •...
  • Page 34: Storage

    5. Maintenance and Storage 5.2 Storage Keep the device in its storage case when not in use. Unplug the air plug from the air jack. Gently fold the air tube into the arm cuff. Note: Do not bend or crease the air tube excessively. Place the monitor and the arm cuff in the storage case.
  • Page 35: Optional Parts

    6. Optional Parts 6.1 Optional Medical Accessories (within the scope of EC Medical Device Directive 93/42/EEC) Arm cuff Small arm cuff AC adapter Arm circumference Arm circumference 22-42 cm 17-22 cm AC ADAPTER-S AC ADAPTER-UK HEM-RML31 CS2 Small Cuff (60240HW5SW) (60240H7000SW) (HEM-CS24) Using the Optional AC Adapter...
  • Page 36: Specifications

    7. Specifications Product category Electronic Sphygmomanometers Product description Automatic Upper Arm Blood Pressure Monitor Model (code) MIT5s Connect (HEM-7280T-E) Display LCD digital display Measurement method Oscillometric method Transmission method Bluetooth Version 4.0 (Low Energy support) Wireless communication Frequency range: 2.4GHz (2400 - 2483.5 MHz) Modulation: GFSK Effective radiated power: <20 dBm...
  • Page 37 Smart word mark and logos are registered trademarks owned by Bluetooth SIG, ® Inc. and any use of such marks by OMRON HEALTHCARE Co., Ltd. is under license. Other trademarks and trade names are those of their respective owners. Apple and the Apple logo are trademarks of Apple Inc., registered in the U.S. and other countries.
  • Page 38 7. Specifications Hereby, OMRON HEALTHCARE Co., Ltd., declares that the radio equipment type MIT5s connect (HEM-7280T-E) is in compliance with Directive 2014/53/EU. The full text of the EU declaration of conformity is available at the following internet address: www.omron-healthcare.com Important information regarding Electro Magnetic Compatibility (EMC) With the increased number of electronic devices such as PC’s and mobile (cellular) telephones, medical devices in...
  • Page 39: Warranty

    This product is guaranteed by OMRON for a period of 2 years after the date of purchase. The proper construction, workmanship and materials of this product is guaranteed by OMRON. During this period of guarantee OMRON will, without charge for labour or parts, repair or replace the defect product or any defective parts.
  • Page 40: Some Useful Information About Blood Pressure

    9. Some Useful Information about Blood Pressure What is Blood Pressure? Blood pressure is a measure of the force of blood flowing against the walls of the arteries. Arterial blood pressure is constantly changing during the course of the heart’s cycle. The highest pressure in the cycle is called the Systolic Blood Pressure;...
  • Page 43 Package includes: Inhoud van de verpakking: Contenu de l’emballage : Комплект поставки содержит: Packungsinhalt: Paketin içindekiler: :‫ﺗﺣﺗوي اﻟﻌﺑوة ﻋﻠﻰ‬ Il pacchetto include: El paquete incluye:...
  • Page 44 No.28 VSIP II, Street 2, Vietnam-Singapore Industrial Park II, Binh Duong Industry-Services-Urban Complex, Hoa Phu Ward, Thu Dau Mot City, Binh Duong Province, Vietnam Subsidiaries Empresas filiales Succursales Dochterondernemingen Niederlassungen Филиалы Consociate Yan Kuruluşlar...