MODEL 8559A
5-18.
CALIBRATED SWEEP TIME ADJUSTMENT (Cont'd)
1.
Set equipment controls as follows:
If an HP 853A Spectrum Analyzer Display mainframe is used, and a sweep
time faster than 10 msec is selected, an error message will appear on the
analyzer's CRT and the analyzer will go into mixed mode.
Spectrum Analyzer:
TUNING
FREQ SPAN/DIV
RESOLUTION BW
INPUTATTEN
REFERENCE LEVEL
REFLEVELFINE
Amplitude Scale
SWEEP TIME/DIV
SWEEP TRIGGER
ALTIF
SIGIDENT
BLCLIP
VIDEOFILTER
50 MHz Universal Counter:

FUNCTION

SAMPLE RATE
A 50 MHz INPUT
B 10 MHz INPUT
2. Connect equipment as shown in Figure 5-2.
3. Adjust counter SENSITIVITY controls (both channels) as necessary until counter triggers and indicates a
time interval of approximately 10.00 ms.
4.
Adjust 1 ms potentiometer A9R10 for a time interval indication of 10.00
ADJUSTMENTS
NOTE
ADJUSTMENTS
F (full)
Optimum, coupled
10 dB/DIV
FREE RUN
OFF
OFF
OFF
OFF
TI. A to B
Full counterclockwise
(rising edge)