ADJUSTMENTS
5-18.
CALIBRATED SWEEP TIME ADJUSTMENT (Cont'd)
In early instruments, A9R13 is labeled "2 ms." The adjustment of A9R13,
however, should be performed with SWEEP TlMUDlV set at 5 ms.
5 .
spectrum analyzer SWEEP TIME/DIV control to 5 msec. Readjust counter SENSITIVITY controls
Set
as necessary and adjust 5 ms potentiometer A9R13 for a time interval indication of
6. Check time interval for each SWEEP TIME/DIV control setting listed in Table 5-5. Readjust 1 ms potenti-
ometer A9R10 and 5 ms potentiometer A9R13 as necessary if test limits are exceeded.
SWEEP TIMEIDIV
ADJUSTMENTS
NOTE
TABLE 5-5. CALIBRATED SWEEP TIMETEST LIMITS
Setting
2 ms
5 ms
2 0 ms
5 0 ms
50.00
Sweep Time (ms)
MODEL 8559A
4.00
ms.