ADJUSTMENTS
5-25.
FIRST CONVERTER ADJUSTMENTS
REFERENCE:
A3, A4, AS, A6, and A7 Schematics
DESCRIPTION:
The First LO (A6 YTO) is adjusted by monitoring the YTO output at the RF input connector (LO emission) and
the tuning voltage (TUNE) output of the A7 Frequency Control board, and adjusting the YTO low-end fre-
quency for 3 GHz at O V tuning voltage and 6 GHz at
D l SPLAY
n
EXTENDER
CABLE
ASSEMBLY
EQUIPMENT:
Digital Voltmeter
Comb Generator
Cable, BNC (m) to Banana Plug
Extender Cable Assembly
Adapter, BNC (f) to Alligator Clips
Cable Assembly, RG-214/U, Type N Connectors
PROCEDURE:
1. Allow one-half hour warmup time of equipment with spectrum analyzer connected to mainframe with
extender cable.
ADJUSTMENTS
-
CABLE
.
LOCATED THROUGH
HOLE I N S I D E
FRAME HERE
FIGURE 5-1 1. FIRST CONVERTER ADJUSTMENTS TEST SETUP
MICROWAVE
COUNTER
ASSEMBLY
TUNE
CABLE
MODEL 8559A
QENERATOR
OUTPUT
H P 3456A
H P 8406A
H P 101 1 1A
H P SO6O-0303
H P 8120-1292
H P 1 1500A

Advertisement