HP 8559A Operation And Service Manual page 101

Spectrum analyzer
Hide thumbs Also See for 8559A:
Table of Contents

Advertisement

MODEL 8559A
Para.
No.
Sweep Time Accuracy
7. Sweep Time, 2 psecldiv
8. Sweep Time, 5 psec/div
Sweep Time, 10 psecldiv
Sweep Time, 20 psecldiv
Sweep Time, 50 psecldiv
Sweep Time, . l msec/div
Sweep Time, .2 msecldiv
Sweep Time, .5 msec/div
Sweep Time, 1 msecldiv
Sweep Time, 2 msec/div
Sweep Time, 5 msec/div
Sweep Time, 10 msecldiv
12. Sweep Time, 20 msec/div
Sweep Time, 50 msec/div
Sweep Time, .l secldiv
Sweep Time, .2 secldiv
Sweep Time, .5 secldiv
Sweep Time, 1 secldiv
Sweep Time, 5 sec/div
Sweep Time, 10 secldiv
Calibrator Output Accuracy
1. Calibrator Output Frequency
2. Calibrator Output Power (50R)
Display Fidelity
9. Log Maximum Cumulative Error
10. Log Incremental Error
16. Percent Linear Error
(Percent of Reference Level)
TABLE 4-18. PERFORMANCE TEST RECORD (4 OF 4)
Test Description
Results
Actual
Min.
1.80 psec
4.50 psec
9 .OO psec
45 . O O psec
90.00 psec
180 psec
450 psec
900 psec
4500 psec
9.00 msec
0.180 sec
0.450 sec
0.90 sec
1.80 sec
4.50 sec
9.0 sec
40.0 sec
80.0 sec
34.600 MHz
-10.3 dBm
-1.5 dB
-0.1 dB/dB
PERFORMANCE TESTS
Max.
2.20 psec
5.50 psec
11 .OO psec
22.00 psec
55.00 psec
110 psec
220 psec
550 psec
2200 psec
5500 psec
1 1 .OO msec
0.220 sec
0.550 sec
1.10 sec
2.20 sec
5.50 sec
60 sec
120 sec
35.400 MHz
-9.7 dBm
3%

Advertisement