ADJUSTMENTS
5-19. LOG AMPLIFIER LOG AND LINEAR ADJUSTMENTS (Cont'd)
PROCEDURE:
1. Set equipment controls as follows:
Spectrum Analyzer:
FREQUENCY BAND GHz
TUNING
RESOLUTION BW
INPUTATTEN
REFERENCELEVEL
REFLEVELFINE
Amplitude Scale
SWEEP TIME/DIV
SWEEPTRIGGER
ALTIF
BLCLIP
VIDEOFILTER
Signal Generator:
COUNTERMODE
AM
FM
FREQUENCY TUNE
RF
OUTPUTLEVEL
2. Set 1-dB step attenuator to 10 dB and 10-dB step attenuator to 0 dB. Remove AlOWl (blue cable) from
The HP 355C 10 dB attenuation is included to compensate for 10 dB of gain
on Step Gain Assembly A12 with the TEST-NORM switch in TEST.
3. Set TEST-NORM switch on Step Gain Assembly A12 to TEST position. Adjust signal generator FRE-
QUENCY TUNE control for maximum signal amplitude on display with 10-dB step attenuator set to
(reduce signal generator OUTPUT LEVEL control setting as necessary to bring signal on-screen).
ADJUSTMENTS
NOTE
MODEL 8559A
3
-
300 kHz, uncoupled
LIN
AUTO
OFF
OFF
OFF
OFF
INT
OFF
OFF
321.4 MHz
ON
approx. -28dBm
dB