Temperature Out Of Range - Agilent Technologies 1100 Series Reference Manual

Quaternary pump
Hide thumbs Also See for 1100 Series:
Table of Contents

Advertisement

Probable Causes
Suggested Actions
Troubleshooting and Test Functions

Temperature Out of Range

Temperature Out of Range
The temperature sensor readings in the motor-drive circuit are out of range.
The values supplied to the ADC by the hybrid sensors must be between 0.5 V
and 4.3 V. If the values are outside this range, the error message is generated.
• Defective LPM board.
❏ Exchange the LPM board.
72