Startup And Operation Of Hazardous Area Oxymitter 4000 With Loi; Power Up; Start Up Oxymitter 4000 Calibration; Figure 6-1. Startup Display - Emerson Rosemount Analytical Oxymitter 4000 Instruction Manual

Hazardous area oxygen transmitter
Table of Contents

Advertisement

Hazardous Area Oxymitter 4000
OF HAZARDOUS AREA OXYMITTER 4000 WITH LOI
6-1

POWER UP

a. Startup Display
When power is applied to the probe, the cell
heater turns on. It takes approximately one
half hour for the cell to heat to operating
temperature. This condition is indicated by a
"warm up" display on the LOI (Figure 6-1).
This message will continue to display until
the cell is up to operating temperature.
b. Operating Display
The normal operating display is the % O
concentration. The "normal" display is
shown in Figure 6-2.
c. Error
If there is an error condition at startup, an
alarm message will be displayed. Refer to
Section 8, TROUBLESHOOTING, to deter-
mine the cause of the error. Clear the error,
cycle power, and the % O
return.
d. LOI
The Local Operator Interface can be used to
change the software and alarm settings, to
adjust the high and low gas settings, and to
initiate the calibration sequence. Refer to
the LOI menu (Figure 6-4).
e. Reference Air
Ensure the reference air, if used, is set to
56.6 l/hr (2 scfh).
6-2
START UP OXYMITTER 4000
CALIBRATION
Refer to Section 9, MAINTENANCE AND
SERVICE, for calibration instructions.
Rosemount Analytical Inc. A Division of Emerson Process Management
SECTION 6
STARTUP AND OPERATION
2
display should
2
Instruction Manual
IB-106-340C Rev. 4.1
O2:
0.00%
warm up
367dgC

Figure 6-1. Startup Display

O2:
2.59%
normal

Figure 6-2. Normal Display

Startup and Operation with LOI
July 2004
LK
37260035
6
LK
37260036
6-1