Maintenance and Service
ALARMS
LED
FLASHES
1
2
HEATER T/C
3
4
1
2
3
HEATER
4
5
1
3
O2 CELL
4
1
CALIBRATION
2
3
NOT USED

Figure 9-2. Inside Right Cover

Push the CAL key. The CALI-
BRATION RECOMMENDED
LED will turn off and the CAL
LED will flash continuously.
The Hazardous Area Oxymit-
ter 4000 can be configured
so that the 4-20 mA signal
will hold the last value. The
default condition is for the
output to track. A flashing LED
indicates that the Hazardous
Area Oxymitter 4000 is ready
to accept the first calibration
gas.
Apply the first calibration gas.
(Electronics will abort the cali-
bration if step 4 is not done
within 30 minutes).
Push the CAL key; the CAL
LED will be on solid. A timer is
activated to allow the calibra-
tion gas adequate time to flow
(default time of five minutes).
When the timer times out, the
Hazardous Area Oxymitter 4000
MANUAL
CALIBRATION
*
PLACE CONTROL LOOP
IN MANUAL
STATUS
*
IF CAL LED ON
OPEN
GO TO STEP 2
SHORTED
PUSH CAL
1
CAL LED ON
REVERSED
PUSH CAL
A/D COMM
2
CAL LED FLASH
ERROR
OPEN
APPLY TG1
3
HIGH HIGH
PUSH CAL
TEMP
CAL LED ON SOLID
4
HIGH CASE
WAIT FOR FLASH
TEMP
5
REMOVE TG1 & APPLY TG2
LOW TEMP
PUSH CAL
HIGH TEMP
CAL LED ON SOLID
OPEN
WAIT FOR FLASH
6
BAD
2 FLASH-VALID CAL
3 FLASH-INVALID CAL
EPROM
CORRUPT
7
REMOVE TG2
INVALID SLOPE
PUSH CAL
INVALID
CAL LED ON FOR
CONSTANT
8
PURGE TIME
LAST CAL
CAL LED OFF
FAILED
SW2 DIP SWITCH
HART
LOCAL
0-10%
0-25%
4mA
20mA
NOT USED
Rosemount Analytical Inc. A Division of Emerson Process Management
29770005
Hazardous Area Oxymitter
4000 has taken the readings
using the first calibration gas
and the CAL LED will flash
continuously. The flashing
indicates the Hazardous Area
Oxymitter 4000 is ready to
take readings using the sec-
ond calibration gas.
5
Remove the first calibration
gas and apply the second
calibration gas. (Electronics
will abort the calibration if
step 6 is not done within 30
minutes).
6
Push the CAL key; the CAL
LED will be on solid. The timer
is activated for the second
calibration gas flow. When the
timer times out, the CAL LED
will flash a 2 pattern flash or a
3 pattern flash (2 pattern flash
equals a valid calibration, 3
pattern flash equals an invalid