LOGIC I/O CONFIGURATIONS

The Hazardous Area Oxymitter 4000 logic I/O
output can be configured for ten different modes
through HART/AMS. The factory default condi-
tion is Mode 5. A list of possible configurations
appear in Table 7-1.
The Unit Alarm configuration available for
Modes 1, 3, 5, and 7 refers to the diagnostic
alarm faults in Table 8-1.

Table 7-1. Logic I/O Configuration (as set at HART/AMS or LOI)

Mode
0
The unit is not configured for any alarm condition.
1
The unit is configured for a Unit Alarm.
2
The unit is configured for Low O
3
The unit is configured for both a Unit Alarm and Low O
4
The unit is configured for a High AC Impedance/CALIBRATION
RECOMMENDED.
5*
The unit is configured for both a Unit Alarm and a High AC Imped-
ance/CALIBRATION RECOMMENDED.
6
The unit is configured for both a Low O
RECOMMENDED.
7
The unit is configured for a Unit Alarm, a Low O
Impedance/CALIBRATION RECOMMENDED.
8**
The unit is configured for a calibration handshake with an IMPS 4000 or SPS
4000. CALIBRATION RECOMMENDED will initiate the calibration cycle.
9
The unit is configured for a calibration handshake. CALIBRATION RECOM-
MENDED will not initiate the calibration cycle with an IMPS 4000 or SPS 4000.
*The default condition for a Hazardous Area Oxymitter 4000 without an IMPS 4000 or SPS 4000.
**The default condition for a Hazardous Area Oxymitter 4000 with an IMPS 4000 or SPS 4000.
7-4
HART/AMS
Hazardous Area Oxymitter 4000
7-6
HART/AMS MENU TREE FOR HAZARDOUS
AREA OXYMITTER 4000 APPLICATIONS
This section consists of a menu tree for the
HART Communicator. This menu is specific to
Hazardous Area Oxymitter 4000 applications.
Configuration
.
2
.
2
and High AC Impedance/CALIBRATION
2
, and a High AC
2