Hazardous Area Oxymitter 4000
DIAGNOSTIC
ALARMS
CALIBRATION RECOMMENDED
TEST
POINTS
INC
INC
HIGH
LOW
GAS
GAS
DEC
DEC
KEYPAD
Alarms
O2 Heater Open
LOI
Figure 8-7. Fault 5, Open Heater
Rosemount Analytical Inc. A Division of Emerson Process Management
HEATER T/C
HEATER
SW2
O2 CELL
CALIBRATION
J1
O2 CELL mV +
TP1
O2 CELL mv -
TP2
HEATER T/C +
TP3
HEATER T/C -
TP4
CAL
TEST GAS +
TP5
PROCESS -
TP6
% O2
37260023
Instruction Manual
e. Fault 5, Open Heater
Figure 8-7 shows the electronic assembly
for a Hazardous Area Oxymitter 4000 with
a membrane keypad (upper view) and a
Hazardous Area Oxymitter 4000 with an
LOI (lower view).
Membrane Keypad. When Fault 5 is
detected, the HEATER LED flashes once,
pauses for three seconds, and repeats.
1. Remove power.
2. Remove the electronic assembly per
paragraph 9-4c, Electronic Assembly
Replacement.
3. Using a multimeter, measure the re-
sistance across the terminals of heater
connector, J8.
4. The measurement should be approxi-
mately 72 ohms. If the heater is open,
see paragraph 9-4g, Heater Strut
Replacement.
LOI. When Fault 5 is detected, the LOI dis-
plays the "O2 Heater Open" message.
1. Remove power. Unscrew and remove
the LOI module from the electronic as-
sembly.
2. Perform the diagnostic steps 2 through
4 shown for the membrane keypad.
IB-106-340C Rev. 4.1
July 2004
Troubleshooting
8-9
8