o. Fault 15, Last Calibration Failed
Figure 8-17 shows the electronic assembly
for a Hazardous Area Oxymitter 4000 with
a membrane keypad (upper view) and a
Hazardous Area Oxymitter 4000 with an
LOI (lower view).
Membrane Keypad. When Fault 15 is
detected, the CALIBRATION LED flashes
three times, pauses for three seconds, and
repeats.
1. The last calibration failed alarm acti-
vates when the slope and constant
values calculated are out of range and
the unit reverts to using the previous
calibration values.
2. The cell should be replaced. See para-
graph 9-4h, Cell Replacement, for cell
replacement instructions.
LOI. When Fault 15 is detected, the LOI
displays the "Calib Failed" message. Refer
to the comments in paragraphs 1 and 2
above.
IB-106-340C Rev. 4.1
July 2004
Troubleshooting
8-19