Immunity To Rf Wireless Communication Equipment - AGFA DX-D 300 User Manual

Hide thumbs Also See for DX-D 300:
Table of Contents

Advertisement

Note: The ISM (industrial, scientific and medical) bands between 0.15 MHz
and 80 MHz are 6.765 MHz to 6.795 MHz; 13.553 MHz to 13.567 MHz;
26.957 MHz to 27.283 MHz; and 40.66 MHz to 40.70 MHz. The amateur ra-
dio bands between 0.15 MHz and 80 MHz are 1.8 MHz to 2.0 MHz; 3.5 MHz
to 4.0 MHz; 5.3 MHz to 5.4 MHz; 7 MHz to 7.3 MHz; 10.1 MHz to 10.15
MHz; 14 MHz to 14.2 MHz; 18.07 MHz to 18.17 MHz; 21.0 MHz to 21.4
MHz; 24.89 MHz to 24.99 MHz; 28.0 MHz to 29.7 MHz; and 50.0 MHz to
54.0 MHz.

Immunity to RF wireless communication equipment

ISM Band
(MHz)
300-390
430-470
704-787
800-960
1700-1990
2400-2570
5100-5800
DX-D 300 | Remarks for HF-emission and immunity | 83
(80% AM at
(80% AM
1 kHz)
at 1 kHz)
Modulation
Pulse modulation 18 Hz
FM
± 5 kHz deviation
1 kHz sine
Pulse modulation
217 Hz
Pulse modulation
18 Hz
Pulse modulation
217 Hz
Pulse modulation
217 Hz
Pulse modulation
217 Hz
Distance
Immunity test level
(m)
(V/m)
0.3
27
0.3
28
0.3
9
0.3
28
0.3
28
0.3
28
0.3
9